ARRAY
SOLUTIONS
2611 North Beltline Road
Suite 109
Sunnyvale, Texas 75182
USA
ã Dec 22, 2012
Table of Contents
Quick Start..................................................................... 5
Menu Chart ................................................................... 9
Calibration..................................................................... 12
Colors and Styles........................................................... 16
Command Buttons......................................................... 18
Point Data...................................................................... 18
File Commands.............................................................. 23
Function commands....................................................... 25
Stub Tuning.....................................................................27
Calibration commands....................................................30
Setup commands ........................................................... 31
Band commands............................................................. 34
Marker commands ......................................................... 35
Utilities .......................................................................... 39
Help commands.............................................................. 40
Data Display .................................................................. 41
Custom Calibration……………………………………. 43
Data Referred to Antenna…………..……….………… 48
Line Extension………………………………………… 50
Rulers and Offsets........................................................... 52
Scan Differences............................................................. 55
Smith charts..................................................................... 59
DSP Filters....................................................................... 61
TDR (Time Domain Reflectometry)............................... 65
TDR Operation................................................................ 72
AIM4170 Principles of Operation................................... 75
Measurement of Crystal Parameters............................... 79
Frequency Source............................................................ 83
Inductance/Capacitance Measurements.......................... 85
Appendix 1 – Specifications…….……….…….………. 88
Appendix 2 – Complex Numbers…………..….....….…. 91
Appendix 3 – Hot keys………………………….……… 95
Appendix 4 – RS232/USB Operation……………….….. 96
Appendix 5 – Scan Data File Format……………….…... 101
Appendix 6 – Battery Operation…..………………….… 104
Appendix 7 – Saving Screen Shots……………………... 106
Appendix 8 – Configuration File……………………….. 107
Appendix 9 – Component Test Fixture…………………. 113
Appendix 10 - External Control….…….………………. 115
Appendix 11 - Tone and Voice Output… ……………... 117
Appendix 12 – Band Skipping…………….……………. 119
Appendix 13 - Expansion Port for AIMuhf...................... 120
Appendix 14 - Sync output .............................................. 122
Terms and Conditions….…………….….…….……....... 126
Update
History Latest program: www.w5big.com/prog_update.htm
Sept 27 2012
The stub tuning function has been generalized for any fraction of a wavelength
Aug 28 2012
Major changes to the Point Data function to aid in production testing.
May 21 2012
Changed the word Theta to Phase. The calculations and the data are the same as before. Some of the pictures in the manual may show Theta but the meaning is identical: Phase=arctangent(Xs/Rs).
SWR scale can be offset by one vertical division to improve readability in certain cases.
Smith chart trace can be displayed in varying colors across the frequency scan range.
Feb 2012
Added DSP filters for all models of the AIM.
Dec 19, 2011: version 852
DLL functions have been implemented for controlling the AIM with a user written program.
Data logging can be turned on and off while in the Point Data Tune mode.
Oct 10, 2011: version 846
New Difference of Scans function.
TDR range has been increased to 2000 feet.
Aug 2, 2011: version 845A
Scanning rate can change dynamically for faster scanning of a large frequency range.
July 21, 2011: version 845
Plot inductance or capacitance vs. frequency.
Optional labels for traces.
TDR files can be viewed in demo mode.
Sync output documentation added in Appendix 14.
April 30, 2011: version 840
Added TDR function.
Both swr rulers are displayed on the Smith charts.
March 8, 2011: version 835
Display cursor data and the
differences for two points.
The recycle depth can be set using
the Setup Menu.
The crystal analysis function now
goes down to 25KHz.
Information on the usb interface as been added to Appendix 4.
A data Smoothing function has been added to replace Trend.
An offset can be applied to the plotted parameters.
Two rulers can be displayed for plotted parameters.
Overlay the last two graphs
by pressing CNTL+Up Arrow.
The present state of all functions can be saved in a user
specified file.
Quick Start
Loading the Software:
The AIMuhf, the AIM and PowerAIM
have the same basic features. The
primary difference is the maximum frequency.
The maximum frequency capability of the instrument is determined
automatically when the program starts.
In this manual, references to the AIM also apply to the AIMuhf and
PowerAIM.
The AIM software does not require a formal installation procedure. It does not interfere with any other programs or the registry on your computer. Several versions of the program can be resident at the same time.
1) Create a folder or a subfolder on any convenient hard drive.
For example, “C:\AntennaAnalyzer”.
2) Download the latest program:
AIM Program Update. (http://www.w5big.com/prog_update.htm) This program works with the AIMuhf, the AIM4170 and the PowerAIM 120.
3) Unzip all the files in the folder you just created for the analyzer software and documentation.
4) The file labeled “AIM_xxx.exe” is the executable file. It is ready to run without going through an installation process. When later versions of the program are released, the number “_xxx” will be different. All versions of the analyzer program can reside in the same folder at the same time, although separate folders can be used to keep the files organized. The older programs with the lower numbers will not interfere with the newest version, so they do not have to be deleted. If there is any question about the performance of a new version, you can switch back immediately to an earlier version for a comparison.
If you want to make a shortcut icon for your desktop, right click on the AIM_xxx.exe file and select “create shortcut” from the dropdown menu. Drag the shortcut to your desktop or task bar. Right click on the icon to bring up the renaming option.
If the files are loaded from a CD, they may be flagged as “read-only”. In order for the calibration and initialization data files to be updated properly, the “read-only” flag must be cleared. Highlight all the files in the folder by pressing control-A and then right-click to see the properties of the files. If the “read-only” property box has a checkmark, uncheck it and then click “Apply”. This will remove the read-only flag on all the highlighted files. None of the files in the folder need to be read-only.
For operating tips or to post your own ideas check: http://www.w5big.com/forum.htm
HARDWARE CONNECTIONS:
Plug in the DC power supply (8 to 12V at 500 ma recommended) and insert the connector into the jack on the rear panel of the analyzer. Note: if a power supply is included with your AIM, it is for 120VAC only. [The Global Power Supply (120-240V) is available as an option when ordering the AIM.]
Press the power switch. The Green LED will blink a few times to indicate the version of software and then remain on continuously. The Red LED is on only when a measurement is in progress.
To turn off the power, press the power switch again. If the analyzer does not receive a command from the PC for 10 minutes, it will power down automatically if it is in the AutoPwrOff mode. The AutoPwrOff mode can be turned on/off with a menu selection under the Setup tab at the top of the screen.
When using the
AIM with a new antenna system, check the AC and DC voltage between the
antenna ground and the ground used for the AIM and the PC. This
voltage should be less than 1V. A
balanced antenna should have a DC connection to ground through a balun or RF
choke on one side (or both sides). Of
course there should always be a large value resistor (or balun)
connecting both leads of any antenna to a ground path in order to drain off
static electricity. This protects the
AIM and any capacitors that are used in the antenna tuner.
NOTE: Before connecting a transmission line to the
input of the AIM, be sure to momentarily short its pins together to drain off
any static charge that may be present.
Also, be sure there is no DC voltage on the antenna. If there is
DC, use a blocking capacitor between the AIM and the antenna input.
Antennas and transmission lines
can have enough static charge to damage sensitive electronic equipment. This
can happen even when there is no rainstorm in the area. A strong wind can
generate static charge. So can just flexing a coaxial cable by rolling it up or
unrolling it, even if there is no antenna connected to it.
An antenna or a component to be
measured should not be connected or
disconnected from the analyzer while
a test is in progress. A test is in
progress when the RED LED is on.
Be sure the
maximum input voltage at the DC power connector does not exceed 15 volts. The
minimum input voltage required is 7.0 volts.
NOTE: Low cost power supplies that plug in the wall are usually not regulated and their maximum output when no load is connected may be several volts higher than their rated output. Check the output voltage with no load to make sure it does not exceed 15 volts.
Power supplies that operate on a wide
input voltage range such as 120V to 220V use a switching regulator. Evaluate
the AIM measurement results to see if noise from the power supply may be a
problem. For comparison, you can take
some measurements while using a 9V battery and then with the AC power
supply.
The AIM can be operated on battery power for remote operation with a laptop computer. The current required is about 250ma (400 ma for the AIMuhf) while a measurement is in progress and about 50ma when idle.
Batteries are not included with the AIM but you can make a battery pack using any type of batteries you like. There is room inside the case for a 9V battery and disconnect diodes are included so the battery and the AC power supply will not interfere with each other. There is also a space for an optional resistor to use for trickle charging a battery, if desired. The main power on/off switch controls the battery power too, so the leakage current is less than one microamp when the AIM is turned off. Refer to Appendix 6 and the Application-Help file for more details
When using the AIM to test a mobile antenna on a motor vehicle, it is better to use a separate battery and not the 12V battery in the vehicle. A 9V battery will power the unit for several hours or a small 12V gel cell (sealed lead acid) battery can be used for extended operating periods. This avoids the problem of sneak paths through the ground between the DC power input and the antenna ground connection. It will also help reduce measurement noise if it’s necessary to run the engine while taking data (such as to operate the air conditioner). If it’s essential to get power for the AIM from the vehicle, be sure to put 500 ma fuses in BOTH the +12V lead and the power ground lead. A small voltage drop across the fuses will not affect the AIM since the battery voltage is much more than the required minimum operating voltage. The laptop computer being used should remain floating for the best measurement accuracy.
PC INTERFACE CABLE:
Connect one end of the RS232 cable to the analyzer and the other end to COM1 on your PC. If you want to use a different COMM port, start the AIM program and click on the Setup menu at the top of the screen. Then click on Enter Comm Port and enter the port number that you want to use. This number will be saved in the setup file called AIM_xxx.ini.
If you are connecting to the RS232
port with a USB adapter, (or directly to the USB port with an AIMuhf) you will
need the appropriate driver. The
preferred USB adapters use a chip made by FTDI. The drivers for these adapters
can be found here:
http://www.ftdichip.com/Drivers/VCP.htm
For more information on the USB interface, see Appendix 14 of this manual and this document: http://www.w5big.com/PC_USB_INTERFACE.pdf
To set the comm click on Setup -> Enter Comm Port.
The AIM program has been used with Windows versions 95, 98, 2000, XP, Vista and Win 7. It has also been used with MAC’s.
Be sure the configuration file is loaded:
Click Files -> Load Config File and select the file called: _AIM_config.cfg that is in the same folder with the AIM exe file.
Larger version of this image: www.w5big.com/AIM_menu.png
Initial Operation:
Launch the AIM_xxx.exe program. You will see a graph similar to this:
This is the last scan that was taken before the program was
closed. Each time the program starts, it restores the previous scan.
When the AIM is present, it must be turned on before the PC program is launched.
If the PC program starts when the AIM is not connected and powered up, the program will automatically enter the DEMO mode. In demo mode you can view previously saved scan files.
After calibrating, leave the calibration resistor attached to the RF connector and click the SCAN button in the lower left corner of the screen.
The Red LED will come on while the scan is in progress. A blue bar will move across the top of the graph as the scan progresses.
You will see a set of traces (similar to the picture below). The actual values are not important, but this shows the AIM is functioning.
Scan using a 200 ohm resistor
for a test load.
The size of the graph can be specified in the Config File. The ‘max graph width’ and ‘max graph height’ parameters can be changed to make the graph compatible with smaller PC monitors. Graph sizes down to 640x480 (and perhaps smaller) can be used. The ratio of the width to the height is not critical. When using a larger monitor, the graph size can be increased, if desired.
To save screen shots for a document or a web page,
the size of the graph can be adjusted to optimize the screen to fit the
document. This eliminates the need to
rescale the image using a paint program (like Photo Shop), which sometimes
distorts the lines in a graph. A
special config file for this purpose can be saved and then easily recalled when
it is needed.
Calibration
The AIM has no internal
adjustments. There are no trim pots or caps inside. All
the calibration is done with the following software procedure.
Click on the Setup menu and select Calibrate -> Standard Cal.
(The Custom Cal option is used for special applications and
will be discussed later.)
A message box will appear near the center of the screen, as shown below:
Attach the connector with an internal short circuit (this is included with the AIM, it’s labeled “short” and the label is green).
After the short circuit is in place, click Continue.
If the wrong load is attached, an alert is shown and you have the option to retry with the proper load or to stop the cal procedure.
The program will run for a few seconds to take several readings of the short circuit. A number in the upper left corner shows a count down value.
Sometimes when calibrating with a
coax stub or an external filter connected to the AIM RF port, the impedance may
appear to be significantly different from a short circuit and a message will appear asking if this is
a valid condition. If the connections are correct and you really do want to
continue with the special external hardware hooked up, click Continue.
Otherwise, click Cancel or Retry.
After the short circuit has been calibrated, a new prompt appears:
Replace the short circuit with an open connector (red label) and then click Continue.
The program will run for a few seconds to take several readings of the open circuit.
For the best accuracy, especially at high frequencies, use an open connector of the same type that is used for the short circuit. This allows the program to compensate for the stray capacitance of the connector itself. For example, 3 pF of stray capacitance due to a BNC connector represents a reactance of about 1K ohm at 50MHz which will appear to be in parallel with the load. This will seriously affect readings at higher frequencies if it is not properly compensated. Therefore, the three calibration devices (open, short, resistor) should be carefully constructed so they have similar stray capacitance and inductance which will be cancelled out by the calibration procedure.
NOTE: If you are using an adapter,
leave it on the output connector of the AIM when calibrating the open
circuit. In this way, the capacitance of the adapter itself will be
included in the calibration process.
For example, If you want to use a BNC-to-binding post adapter, calibrate the short circuit with a jumper wire across the binding posts. Then remove the wire and calibrate the open circuit condition with the adapter still attached. This procedure cancels the stray capacitance of the adapter and you can get accurate readings of discrete components attached to the binding posts. The cal resistor can be a resistor with leads for this case.
The calibration procedure can also
cancel the effect of a short piece of coax
that is used as a jumper to connect the AIM to the load. In this case, the
open, short and standard resistor are connected to the far end of the
jumper. Good results have been obtained
with as much as two feet (60 cm) of
jumper coax. Longer coaxes
can be calibrated with the Custom Cal procedure (discussed later).
If the
jumper is close to 1/4 of a wavelength long, the calibration data may not be
accurate and a warning message is displayed. You can continue to cal but be
sure to check the scan data for reasonable values under this condition. If you’re sure the data is satisfactory, you
can disable this warning message with a flag in the configuration file.
After taking several readings of the open connector, the program will prompt for the standard resistor with a value between 50 ohms and 500 ohms. This resistor should be mounted on a connector like those used during the open and short calibration steps. The exact value is not critical. Enter the value you measured with a digital ohmmeter and click Continue. If you are using the cal resistor (yellow label) that is included with your AIM, the number printed on it should be entered. If you have a different cal standard, use that value.
After reading the standard resistor, the Red LED goes off and the prompt for a comment appears. This optional comment will be saved with the cal file and can be read later in the Status window. It can also be viewed in the cal file itself with any text editor (such as Notepad). After entering the comment a request for the calibration data file name appears. A separate folder can be used for cal files, if desired.
Enter the name of the calibration file and click “Save”. There is no restriction on the name of the cal data file. Its extension is automatically set to .acal (“analyzer cal”). Any of these cal files can be recalled latter using the menu option “File” à Load Cal File.
Then the cal complete message box is displayed as shown below:
Click OK to continue.
Each time the AIM program starts, the
last calibration file that was used will be read to restore the cal data.
Each AIM has
an ID number associated with it. This ID number is attached to each calibration
file so that cal files will not be used with the wrong AIM. This is important
when several AIM’s may be in use in the same work area. This ID number is placed in the cal file
automatically and no additional user input is required. The AIM’s ID number can be seen by clicking
“Help” -> ”About”. If you don’t see the ID number, your AIM PC program is
not the latest version.
This completes the calibration procedure.
NOTE: Before connecting a transmission line to the
input of the AIM, be sure to momentarily short its pins together to drain off
any static charge that may be present.
Also, be sure there is no DC voltage on the antenna. If there is DC, use
a blocking capacitor between the AIM and the antenna input.
Antennas and transmission lines
can have enough static charge to damage sensitive electronic equipment. This
can happen even when there is no rainstorm in the area. A strong wind can
generate static charge. So can just flexing a coaxial cable by rolling it up or
unrolling it, even if there is no antenna connected to it.
An antenna or a component to be
measured should not be connected or disconnected from the analyzer while a
test is in progress. A test is in
progress when the RED LED is on.
Colors and Styles
The color of each trace corresponds to the color of
the label at the top of the Y-axis. SWR
is RED and this scale is on the left side of the
graph. On the right side the magnitude of the impedance
is the inside scale and the trace is GREEN. The reactance
is in YELLOW, also on the inside scale.
Reactance and impedance use the same scale. Reactance can be positive
(inductive) or negative (capacitive). The phase angle
of the load impedance is plotted in MAGENTA
and this scale is on the outside of the right hand vertical axis. These colors can be changed in the
configuration (*.cfg) file if desired.
The width of the
traces
can be specified in the config file. The default width is “2”. For some
situations, such as presentation slides, a wider line may be desired. A
separate config file can be set up so you can switch between display options
quickly with the Files->Load config
file menu.
Each trace can be
plotted in a different style to make it easier to visualize. This is
particularly helpful when graphs are printed in black and white. The menu
is shown below:
This example
shows some of the styles:
When a new style
is selected, the graph is redrawn to show its effect. The styles are saved in
the *.ini file. The style only applies
to the image on the screen, the raw data is not affected. The traces for a “rescan” are plotted with
plain lines.
A legend showing
the graph styles presently in effect can be displayed by clicking “Legend” on the Graph Styles sub-menu.
This will appear in the lower right corner of the graph. The legend window can be repositioned so it
doesn’t interfere with the plot.
Command Buttons
The most common commands use the buttons along the bottom
edge of the screen:
SCAN (S)– Starts the frequency scan
between the specified limits: Start_freq to the End_freq. (see Limits button
below.) Each time the scan button is
clicked, the graph is cleared and the new scan data replaces the previous data
in memory.
RESCAN (R) – A new scan is started but the graph is not
cleared. This makes it easy to see the before and after effects of changes to
an antenna (or any discrete component being measured). The new data replaces the previous data in
memory but both graphs can be viewed simultaneously. RESCAN can also be used to overlay new data on top of a scan that
was loaded from a data file. See “File -> Load” below.
RECYCLE – Scanning
is repeated over and over until the RECYCLE or the HALT button is clicked. This makes it possible to continuously view
the results while adjusting an antenna or tuning a stub. The resonant frequency is displayed above
the graph and it’s updated after each scan during recycle. The scan limits can be adjusted to narrow
the scan range for a faster update rate.
The number of scans between screen erasures is set
using the Setup Menu item called Recycle Depth. If the depth is zero, the screen is never erased while
recycling. If the depth is set to 5 (the default value), the screen will be
erased after every five scans.
POINT DATA (D) - Measure the impedance of a circuit at a specified frequency. The measurement can be a single-shot reading
or repeated about 2 or 3 times per second while the point data window is open. Click the “Tune” button to start the repetitive
mode. As its name implies, the Tune
mode has been found to be very useful when adjusting an antenna tuner. This avoids stressing the power amplifier
when the antenna is mismatched and it eliminates interference to other people
since the AIM output power is less than 20 microwatts.
You can switch between Tune
mode and Single Shot mode any time by clicking the
buttons in the Point Data window.
While in Tune Mode a tone related to the
SWR can be played through the PC speaker to assist in making adjustments
without watching the monitor. Click the button labeled “Tone On” to start the
tone. The parameters of the tone can be changed using the setup routine called
“Tone” under the Setup tab on the main menu at the top of the
screen.
SWR, Zmag and Phase can be sounded out in
Morse code. See the Setup menu to enable this function. The swr tone or Morse code can be played
over an extension phone or cell phone while making adjustments to an antenna. SWR can also be sounded out using speech wav
files for the set of numbers. These files can be customized for other
languages.
If the Data Log
button is highlighted, the Point Data is written to a data log file. The log option can be turned on and off
any number of times during one session. Only the data read while the data log
button is highlighted will be saved. In
the Single Shot mode, one reading is taken each time the Single Shot button is
clicked or each time the Space Bar is pressed. This makes it easy to data log a
number of components at a particular frequency. The frequency can be changed by clicking the frequency increment
or decrement buttons.
This file can be saved with a user
specified name when the tuning session ends by clicking the Stop/Exit button. The data file has the default extension
*.AIMlog but it can have the extension
*.csv if you like. The log file
can be imported into a program like Excel for
customized processing.
Point Data Window for
AIMuhf
Measured impedance at the AIM RF port = Z = Rs + j*Xs
Rp and j*Xp correspond to the
equivalent parallel circuit values.
Zref
is the reference impedance (typically 50 ohms) used to calculate SWR and the
reflection coefficient, s11. Zref can be any desired
value (including a complex value). It
is entered using the Setup menu
-> Enter Zo Ref
The Fast Log mode gives a faster measurement rate
while data logging, which may be useful when manually adjusting a circuit.
Data Log File Example
LIMITS (L)– When this button is clicked, a dialog box, as shown below, pops up
for entering the start and stop frequencies and the size of the frequency increment
(step) between measurement points.
The start and stop frequencies range from 0.05MHz to 180MHz for the AIM
and 0.05MHz to 1GHz for the AIMuhf.
For example,
to scan the 40-meter band, you might enter 6.9MHz for the Start frequency,
7.4MHz for the Stop frequency and 0.01MHz for the frequency Step (the spacing
between measured points). This would result in a scan of 50 points across the
band. The maximum number of scan points
is 30,000 and the minimum number is 5.
Frequency values can be in KHz instead of MHz if the number is followed by a K or k.
There is an option to begin the X-axis of
the plot at zero rather than at the
Start Freq. This makes the plot easier to read in some cases. The scan will
still begin at the specified Start Freq which cannot be less than the minimum
frequency of the instrument. For example, if you want
to scan from 1MHz to 20MHz, the graph will be numbered starting a 1 and go up
to 26. This makes the numbers on the
X-axis: 1,6,11,16,21,26. However, if you
start at 1MHz, stop at 20MHz and select the “start plot at zero” option, then
the X-axis will be labeled: 0, 5,10,15,20 and the actual scan will be from 1MHz
to 20MHz.
If you enter a number greater than 20 in the Step
box, the scan will use that many data points. (A step size equal to or
greater than 20 is interpreted as a number of points rather than a frequency step in MHz.) If the number in the Step box is followed by a percent sign “%”,
the step size will be that percentage of the total scan width.
Note that when using the cursor to read out numeric
data (discussed in detail later), the displayed values are interpolated between the measured values. Therefore, in some
cases it may be desirable to use a larger frequency step for a faster scan rate. You can
still read the parametric values at intermediate frequencies with the
cursor. The scan rate is about 30 to 60
points per second, depending on the speed of the PC and the value of averaging
that is used.
SCALES - When this button is clicked, a dialog box
pops up for entering the full-scale values for the traces.
If the actual
measured value is off the scale (flat line at the top of the graph), the value
readout by the cursor is still valid since it uses the raw data stored in
memory. For example, you can set the
Zmag scale to 500 ohms in order to see fine details but if the impedance
actually goes up to 1600 ohms at some frequencies, the cursor can still read
the true value and display it in the data window on the right side of the
screen.
There is an option to plot the SWR with swr=1
on the center line (y=0) or shifted up by 1 division. Clicking the Offset entry in the SWR scales column
will alternate between shift or no shift.
The offset does not affect the display when Log SWR is selected, in that
case swr=1 is always on the y=0 axis.
If you would like a full scale value that is not in
the drop-down list for a parameter, you can manually enter any value in the box at the top of the list. For example, if you want to evaluate a 50
ohm resistor, you can enter a full scale value of 60 in order to get better resolution on the graph. (this does
not affect the accuracy of the raw data saved in memory).
Plot Offset and scale factors can be
combined to get better resolution when the parameters don't change much over
the frequency scan range.
If you select a new scale factor, the last data scan
will be replotted using that scale.
COMMENT –
A dialog box pops up for you to enter a comment that will be displayed at the
bottom of the graph. This is very useful for documenting the test conditions.
This comment will appear in a screen capture or a screen print and it will be
saved in the raw data file on disk if this scan is saved. Comments can also
be added to a calibration file before it’s saved.
HALT -
While the scan is in progress, you can stop it by clicking this button. This is
different from the QUIT button (shown below).
QUIT –
This stops the program, saves the setup conditions (limits, scales, etc) and
exits back to the Windows OS. When the
program is launched again, the setup conditions and calibration data will be
restored automatically.
File Commands
Load Graph-- Load a raw data file from a previous scan.
After this file is loaded, its data is just like the original scan. The cursor
can be used to read out the numeric values.
When a data file is being displayed, the name of the
file appears at the top of the graph.
After a graph is loaded, a new scan can be done and
superimposed on top of the old data by clicking Rescan. This is useful for comparing before and
after conditions when adjustments are made or when there may be a long term change in a component or an
antenna. For example, you can see if your antenna is the same today as it was
last week before the windstorm.
When a new graph is loaded from disk with
the same limits as the present graph, it can be overlaid over the existing
graph or it can replace the graph.
Several graphs can be overlaid on each other.
Save Graph --
Save
the raw data for the last scan that was done. If you
clicked the RESCAN button, the data that will be saved is for the rescan. Even
though the earlier scan is being displayed on the graph, its raw data was
replaced in memory by the new data corresponding to last rescan command. The raw data is saved in a file with the
extension .scn. Another file is created with the same name and the extension
.csv. The .csv file has the same data in a format that can be read into a
spreadsheet. The format of this file is shown in Appendix 5.
You can enter an optional comment when
the graph is saved.
Swap Graphs -- Swap back and forth between the present scan and the previous
scan. This function can be disabled by
setting the “autoscansave” flag in the config file to zero (default is
enabled).
Swap can also be done by pressing either the up or the down arrow key.
(Press the CONTROL key and then the arrow key to Overlay the graphs.)
Save Image Bitmap --
Save the current graph image to a file in the .bmp format. If you just want to put the image on the
clipboard to paste it into a document or another program, press Alt-PrintScreen instead using of this menu function.
Then paste it into another application using Control-V. The bitmap image
can be pasted into a paint program or a document, like Microsoft Word.
Print --
Print the graph on the system printer.
Before printing in black and white, you can change the graph display styles to highlight each trace. This option is under the Setup
menu.
Load Calibration -- Load a calibration data file for a particular test setup or adapter.
These files have the .acal extension.
Load Config -- Any of several configuration files can be loaded. A selection window
will show all the files in this folder with the .cfg extension. For example, this can be used to change the
screen size or the colors for different applications. After loading a new config file, close the program and restart
it for the changes to take effect.
Save State -- The present state of the programmable functions can be saved in a file
named by the user. This is the same
data that is normally saved in the *.ini file when exiting from the program. It
allows the user to quickly recall a new set of limits and scale factors for
specific applications. The state file
can also be send to someone else so they can duplicate the test
conditions.
These files have the extension *.stat
Load State -- Load a new state file. Restore the scan limits and scale factors.
DC Power Off -- Remotely turn off the DC power to the AIM. (It cannot be turned on
remotely)
Quit -- Stop
the program and exit.
This is the same as the QUIT button at the bottom of the screen.
Function Commands
Distance to Fault --Measure the distance to a cable open or short. This can also be used to
measure the ¼ wave length frequency of
tuning stubs. The impedance of the cable is also measured.
For accurate results, it is important that the cable
be uniform, that is, it should not be made up of more than one type of
cable.
Refer to Antenna -- The impedance readings are transformed to be
equivalent to readings directly at the antenna terminals. This procedure does not require
disconnecting the transmission line from the antenna if you know the parameters
of the coax.
This should not be used if the transmission line is
made up of several different
kinds of line,
such as, coax cable plus ladder line.
For complex transmission lines, the Custom Cal procedure should be used. It’s discussed later in this manual.
Constant Freq -- Output a constant frequency that
can be used as a test signal.
Initial frequency accuracy is +/-2ppm for the AIMuhf and +/-30ppm for the AIM4170. At 1MHz, the output amplitude into 50 ohms
is about 25 to 30mV-rms (-18 dBm) and somewhat less at higher frequencies. The exact frequency can be calibrated using
WWV or a frequency counter. The default
cal frequency is 10 MHz, but any value can be entered. The initial setting has to be within 2 KHz of
the standard used for calibration.
The Frequency Compensation Factor used to correct all
subsequent programmed values can be seen in the Status Window
(Help->Status). The value is always
very close to unity. The
compensation factor is used to adjust the programmed frequency during a scan
for antennas or components.
Band Scan -- Scan a band (particularly the AM broadcast band) to look for strong
signals that may interfere with antenna measurements. The scan start/stop
limits are set using the same LIMITS button that is used for an impedance
scan. The maximum scan range is 10MHz. The maximum frequency step is 2KHz, but a smaller step can be specified when the limits are set.
The recommended maximum amplitude limit for an external signal is
indicated by a red line, which corresponds to approximately 150mv peak. (This
level is not precisely calibrated.) Signals above this limit may result in less
accurate impedance readings. It is not a rigidly fixed limit and in some cases,
impedance readings may still be sufficiently accurate when external signals are
above the red line.
Click the Rescan
button to repeat the band scan.
Example of a scan in the AM broadcast band
Measure Crystal -- Measure the parameters of a quartz crystal automatically.
Details are in a later section. The crystal data can be saved in a file that
can be imported to a spreadsheet program, such as Excel.
The AIMuhf has a temperature compensated
oscillator (TCXO) which makes it much more
accurate than the AIM4170 when measuring crystals. At room temperature, the drift over several hours is typically
less than 1 Hz for the AIMuhf. The
oscillator can be calibrated in the Constant Freq mode by using a
frequency counter or zero beating with WWV. This calibration will then be
applied to all measurements.
Stub Tuning -- Adjust a coax line to be a
specified fraction of a wavelength at a given frequency. The target frequency
and the initial length (in feet or meters) of the coax are entered in the
dialog box. The unit of measure used
for the cable length is given in the config file. This parameter can be overridden by entering a value for
the cable length followed by the letter "f" or "F" for feet and "m" or "M" for meters.
This value for physical length should be accurate because it is used to
calculate the velocity factor. The wavelength can be entered as a decimal
fraction or as degrees. If the number
is larger than 10, it is assumed to be degrees and it's divided by 360. For example, entering "120" means
the target wavelength is 0.333 (i.e. 120 degrees = one-third of a full wave) at
the target frequency.
The program runs in recycle mode and displays the
amount of cable that needs to be cut off ("Excess Length") to obtain the target wavelength at the target frequency. A
vertical red line on the graph highlights the target
frequency. The magenta colored phase plot
crosses the horizontal axis at the quarter wave frequency. As the cable is trimmed, the present
wavelength at the target frequency is displayed at the top of the graph along
with the amount of cable that needs to be trimmed off.
Data at the bottom of the graph shows the cable
impedance and the velocity factor along with the target values for frequency, wavelength and cable length. The Recycle
Depth indicates how often the graph will be erased while in the Recycle
mode. Zero means don't
erase at all. Five means
erase and start over every five scans. The Recycle Depth value is entered with
the Setup menu, click: Setup -> Recycle Depth
Q Measurement – This is used to measure the Q of resonant
circuits. First, do a scan with the limits set to
include the resonant frequency and most of the response on either side of the
resonant frequency. The step size is not critical but Q may change slightly
with different step sizes. Experiment to see what is appropriate for your
application. The resonant frequency does not have to be in the center of the
graph.
Antenna Bandwidth-- The bandwidth of an antenna at an SWR value set by
the SWR ruler is determined. First,
enter the SWR ruler value with the Setup Menu. Then position the cursor near the valley of
the SWR curve and click “Antenna Bandwidth”. The bandwidth and high/low
frequencies will be displayed in a message box. Vertical green lines will highlight the bandwidth region. For a multiband antenna, the bandwidth at
each resonant frequency can be read by moving the cursor to the desired SWR
valley. The cursor does not have to be exactly at the minimum SWR point.
After positioning the cursor near the desired point,
press the left mouse button to freeze the AIM cursor while
moving to the bandwidth selection button on the Function menu.
Line Extension-- see the Line Extension
section of this manual.
Measure Zo - Measure the impedance of an open or shorted coaxial line.
Calibration
Commands
Standard Cal -- Calibrate the AIM using open circuit, short circuit and resistive load
conditions. The calibration data is saved in a disk file that’s read each time
the program is started.
Any number of calibration files can be saved for
different applications.
An optional comment can saved with each
cal file. The comment for the cal file presently in use can be seen in the
Help->Status window.
Custom Cal -- The start, stop and delta frequency
values can be user specified. The
regular short,
open & resistor loads are used. Long transmissions lines of any type can
be calibrated at the far end so impedance data is then referred to the antenna terminals. It can also be
used to calibrate out the effects of filters in the line. This function takes into account the length
of the cable and also the loss.
After a transmission line or filter is
calibrated, the start, stop and step frequencies can be changed as long as they
are within the original custom cal limits.
For the AIMuhf model, the menu shows several
calibration ranges for quick reference. They are custom cal ranges and the
limits shown in the dialog box can be customized to anything you like.
Setup Commands
Plot Parameters -- Select the parameters that are plotted during a scan. The optional
parameters are: SWR or reflection coefficient or Q,
L or C, Return Loss, Impedance Magnitude, Series Load Circuit or Parallel Load
Circuit, Phase (phase angle). The last
scan data will be replotted with the new parameters.
Graph Styles -- A different style for plotting each trace
can be specified to make the traces more distinctive. This is especially helpful when the graph will be printed in
black and white. For examples of the
styles, refer to the section of the manual above called “Colors and Styles”.
A style legend can be displayed in the lower right corner of the
window. This window can be repositioned
if desired.
Plot Offset Values --
The traces can be offset by a specified amount to make it easier to see small
variations around an average value.
For more information see the Rulers and Offsets Section
Ruler 1 and 2 -- Horizontal
rulers can be defined to help visualize where a parameter lies with respect
to user specified limits. Two rulers can be defined for: SWR, Phase,
Zmag, Rseries, Xseries, and Return Loss. These are referred to as Ruler 1 and
Ruler 2. Either one can have the larger
value.
For more information see the Rulers and Offsets Section
Label Traces-- A letter can be displayed at the beginning
of each trace to help distinguish it.
This is very useful for documentation if the graphs are printed in black
and white.
Enter Zo -- Enter the nominal impedance of the transmission line. This
is used to calculate the SWR. This is usually a real number, but a complex
number can be entered with the imaginary part denoted by “i” or “j”.
Average Readings-- Measurement noise can be reduced by averaging several readings at each
point. When this option is selected, “AVG =N” appears in red in the upper right
corner of the graph. N is the number of
readings being averaged and is between 4 and
256.
Smoothing -- Noise can be reduced by averaging a
number of data values over a narrow frequency range. The averaging process does not shift the frequency of a peak or
phase zero crossing.
The magnitude of the peak or valley of a rapidly
changing parameter may be affected. Check the results with and without
smoothing if there is any doubt.
Auto smoothing uses an averaging size corresponding to
2% of the number of scan points. This
value is recalculated when the scan parameters are changed.
Manual smoothing allows the averaging size to be
specified by the user. This is always an odd number between 3 and 31. A value of 0 turns smoothing off.
Smoothing can be enabled or disabled without changing the value of the manual smoothing
parameter.
Recycle Depth -- In the Recycle mode, this is the number
of scans made before the screen is cleared and it starts over. If the value is
zero, the screen will not be cleared and the scans accumulate indefinitely.
Enter Comm Port -- The value of the comm port is entered.
This can be determined by using the Windows Device Manager. There is more
information in the section about setting up the USB interface.
Find Comm Port -- Depending on the operating system, the comm port can sometimes be found
automatically by the computer. (this
function does not work with all versions of Windows.)
Cursor Normally ON -- If this item has a check mark
beside it, the cursor will move with the mouse and it stops when the left
button is pressed. If this item is not checked, the cursor only
moves when the left mouse button is pressed.
Grid Highlight -- Some grid lines can be highlighted to make the display easier to read.
The appearance of this effect depends on the monitor and it may not be
desirable in all cases.
Audio Out -- Options to output the parametric data through the audio channel in the
Point Data mode. This can make it easier to perform adjustments without having
to watch the monitor.
Auto Power Down -- When using battery power, the AIM can be turned off automatically after
about ten minutes of inactivity to conserve power.
Dynamic Scan Rate -- The scan rate can be changed progressively during a scan to speed up
the process. This is useful when scanning from a low frequency to a much higher
frequency. It is similar to a
logarithmic scan.
The frequency steps get larger from left to right, so
the trace may not be as smooth but it is good for getting a quick overview
while making adjustments.
BAND Functions
Highlight -- Highlight the frequency bands, such as the ham bands or AM or FM radio
bands. These frequencies are specified in the configuration file (*.cfg). The
start/stop limits do not have to be the actual amateur band limits. The highlight color can also be specified in
the config file.
The name of the band can also be user specified to be
more descriptive. Refer to comments in the config file. The AM radio config file has examples
showing the names of each station.
Band Skip -- Scan only in the specified bands that are included in the present
start/stop scan range and skip between bands. This makes it possible to scan
several bands for a multiband antenna much more rapidly.
Click on the desired band to set the start/stop scan
limits and the frequency step between
measurement points. These limits can be
changed in the configuration file.
The title of the band can be changed in
the config file. For example, bands can be
assigned for specific radio stations and their call letters can be used for the
title.
Band A..D are additional user specified band limits.
Same as above.
Marker
Commands
Markers can be defined to aid in collecting data at specific frequencies. Position the mouse near the desired point and right click. The marker frequency at the mouse appears in a dialog box. This value can be changed with a keyboard input. The marker index number is assigned automatically. The markers are always sorted by increasing frequency. A number followed by k or K is considered to be in kilohertz, otherwise, it is in megahertz.
Click the Insert button to add a new marker or the Remove button to remove a marker. When removing a marker, the one closest to the mouse position will be removed. The mouse does not have to be positioned exactly on the marker.
As markers are entered, they are displayed as red dotted lines, as shown below.
The marker display can be turned on or off with the marker toggle function (see below).
After a marker is inserted or removed, the set of markers will be resorted in the order of ascending frequency. Up to 20 markers can be entered at one time. Sets of markers can be saved in files and recalled later.
Markers are controlled using the menu items under the Marker tab:
Several options for the markers are shown in the submenu below:
Display Marker Data – open a window to display all the presently defined markers and the data at these frequencies. The data is read once when the window is opened. Click Refresh to take another set of readings. Click Recycle to read the raw data at the marker frequencies repetitively. In the Recycle mode, the horizontal gauge in red shows the SWR graphically on a log scale from 1 to 5. This corresponds to the actual numeric value of SWR displayed in the third column. This feature is very useful when adjusting multiband antennas since the SWR in each band is displayed in real time while adjustments are made. Several markers can be defined for each band to get an idea of the bandwidth. The following image shows marker data for a two-band trap dipole with five markers per band:
The numbers in parentheses in the first column are the marker index numbers. The markers frequencies do not have to be equally spaced.
If the graph presently displayed is from a scan that has been saved, the marker data will be taken from that data in memory and no new data is read from the hardware. In this case, if some of the markers are outside the scan limits, a reminder will pop up indicating that data for some of the markers cannot be read. Also, the refresh and recycle buttons will be grayed out.
Phase Rotation – The marker data can be transformed to correspond to data that would be measured at the end of a transmission line with a specified electrical length or a network with a specified phase shift.
Toggle Markers - turn the marker display on/off without changing the markers in memory.
Clear all Markers – removes all markers from memory.
Save Markers – saves this set
of markers to a user specified file. Any number of marker files can be
saved. These files have the extension *.mrk.
When the marker data is saved using the SAVE button on the marker data window, a second file is also saved in a comma-separated-variables format (csv). This file as the same user specified name and the extension *.mrx. It can be input to a spreadsheet, such as, Excel.
When the AIM program is closed, the present set of markers is saved in the *.ini file.
Recall Markers - recall a set of markers from a file.
Rotate Smith Chart – A second Smith chart can be displayed. There is an option to rotate its data by a angle corresponding to electrical distance along a transmission line or phase shift in a network. The actual geometric rotation of the plot on the chart will be twice the value of the angle that is entered.
Polar Plot – In addition to the Smith chart, several parameters can be plotted in polar format. The parameters are: SWR, Z and s11.
UTILITIES
L-C Matching Networks by WY2U
This program which is available on the web takes the
impedance data (R and X) that is measured at the input end of the transmission
line and calculates the L-C network required to transform it to the correct
impedance (for example, 50 ohms) to match the transmitter. This can be used
when designing a tuner to find the components to match an antenna over the
desired frequency range.
When you click on this link, your computer must be
connected to the internet. You may have to authorize the connection if you have
a firewall program.
ZPlots by
AC6LA
This program uses the data saved in the .csv file (by
"File Save") to plot graphs for presentations. It features plot
parameter selection and zoom capability.
It can plot parameters that were not included in the original AIM plot
and the plot range can be a subrange of the original data.
When you click on this link, your computer must be
connected to the internet. You may have to authorize the connection if you have
a firewall program. You can download
the Zplots program to your computer and
then run it off-line.
Twelfth Wave Matching - an interesting way to join two coaxes that have different impedances
without using a matching section with a special impedance value.
HELP Commands
AIM Manual -- This manual can be displayed as a pdf file.
Applications -- Information about particular
applications with the AIM. This is in html format and can be edited by the user if desired. The file name is “applications_xx.htm”.
Edit Config File -- The config file is opened in a text editor for making
changes to parameters. After saving the file, be sure to close the AIM program
and restart it to load the new config file.
Status -- The status window
shows the value of some variables and the current path names for files. When Custom Cal is being used, the start,
stop, & step parameters are displayed.
Check Program Update -- Click this link when connected to the internet to go directly to the
program update page at w5big.com.
About -- The
present version of the programs in the PC and the controller are
displayed. The url of the W5BIG website
is also displayed.
Data Display
After a scan (or after loading a file from disk), the
mouse can move a cursor along the frequency axis and the numeric data for
several parameters will be displayed continuously in a data window on the right
side of the screen. An example is
shown below:
The light cyan vertical line is the cursor. It moves
with the mouse whenever the mouse pointer is inside the graph area. In this
example, the frequency is 7.4477 MHz. The frequency changes in 1-pixel
increments due to the mouse resolution, so some specific frequencies may not be
displayable. The data is interpolated
between the actual data points that were recorded during the scan. The cursor
can also be moved in small increments using the Left/Right Arrow Keys on the
keyboard.
Normally the
cursor moves whenever the mouse pointer is inside the graph area. The cursor movement can be stopped by
pressing the left mouse button. This makes it easier to select
items on the menu without disturbing the data being displayed on the right side
of the graph. The
action of enabling/disabling the cursor with the left mouse button can be
reversed with an option on the Setup Menu.
Data in the window shows the characteristic impedance,
Zo, has been specified to be 50 ohms. The SWR at 7.4477 MHz is 21.532,
Zmag=35.239 ohms, and Phase= -84.352 degrees.
Parameter values for both a series and a parallel
equivalent circuit are shown as Rs, Ls (series circuit) and Rp, Lp (parallel circuit). Note that when the phase angle, Phase, is negative,
the equivalent components, Ls and Lp, change to Cs and Cp automatically.
Cursor Data
The data at two frequencies can be displayed along
with their differences by pressing a numeric 1
for the first point, moving the cursor to the second frequency and then
pressing 2 for the second point. This action can be repeated by moving to
other points and pressing 1 or 2. Data in the small window will be updated each time. The two selected frequencies are indicated
by blue vertical lines on the graph.
Close the data window by clicking the x
in the upper right corner.
Information on selecting markers for up to twenty frequencies is on a following page.
At the top of the graph in the main window, up to
five resonant frequencies of the antenna are displayed. These are the
frequencies where the phase angle passes through zero.
The names of the calibration and configuration files being used
are shown in the lower right corner of the data window. The color used to
display these file names can be selected in the configuration file.
Custom
Calibration
This specialized calibration procedure is useful for situations
involving filters or long cables. For
most routine work with physically small sockets, adapters, or short cables, the
Standard Calibration procedure can be
used.
This calibration technique allows the measurement reference point to be
moved from the RF connector on the front panel to the end of a transmission
line and/or a filter. This cancels the
effect of complex transmission lines and filters so the data at the antenna can
be determined more accurately. The
custom cal routine can be used to calibrate long transmission lines or a
relatively short line that may be used for an interconnection in a lab
setup. This calibration process may
take longer when a small delta-freq is used, but the actual scan rate is
essentially the same using either standard or custom cal data.
The user specifies the start, stop and delta frequency values for the calibration. Typically the delta frequency will be 1MHz
or less to take advantage of the improved accuracy when a complex circuit is
connected between the RF connector and the measurement point. Note: the standard calibration procedure
takes cal data at 1MHz increments.
The custom calibration procedure for the AIMuhf can be faster than the standard cal by specifying only the frequency range of interest rather than calibrating over the entire operating range of the instrument.
The AIMuhf has some predefined cal limits that will be automatically
entered into the custom cal dialog.
These values can be modified for specific applications.
The open, short
and resistor loads are used. The resistor value is not critical as long
as it is accurately known. When
adapters are used, be sure to be consistent so that all three loads have the
same capacitance and inductance. For
the open cal load, be sure to use an actual open connector. If the open cal
load is not attached to the test connector, the stray capacity will not be
properly calibrated. This can
represent an error several picofarads.
Click Calibrate -> Custom Cal and the parameter entry dialog box will open.
By default, data entries are in MHz but if a number is followed by a
K or
k then the number will be
interpreted as KHz.
For AIMuhf there are several predefined calibration bands shown
on the menu. These ranges are the same as Custom Cal and the values can be
changed to anything desired. The menu
value are just a shortcut for filling in the dialog box before doing a
custom cal. The three values on each
line are the Start - Stop and (Delta Freq) values.
After the start, stop and delta limits are entered, there will be
prompts to attach the short, open and resistor loads. After the calibration is complete, there is a prompt for an
optional comment to be included with the cal data file. Then the name of the
data file is entered. The name can be anything, but you may want to make the
custom cal files distinctive from the standard cal files so it’s easy to tell
them apart. After calibration, a scan
is done automatically (without plotting the data) to see if the custom cal data
is likely to be accurate enough for this particular setup. When the external circuit has rapid changes
in phase (such as a filter), the custom delta freq has to be small enough to
include data during the phase transitions. In some cases it may be necessary to
use a delta as small as 1 KHz. Up to 30,000 cal points can be used.
If a warning message is
displayed, the cal data may still be ok in the frequency range of
interest. The overall results should be
evaluated with known loads, such as resistors with good RF characteristics.
When a custom cal file is being used, the words “Custom Cal” will appear on the graph. The scan start/stop and delta
limits can be changed without having to do
the custom cal again.
The comment that is included in the current cal file can be viewed by
clicking Help -> Status. The status window
also shows the start, stop and delta parameters used for this custom cal
file. If a custom cal file is not
being used, these data do not appear in the status window.
Custom Cal
Examples:
The effectiveness of the custom cal procedure can be seen in the
following examples. For this series of tests, a transmission line made up of
two different kinds of coax was used:
The high pass filter cuts off below 1.8MHz.
It’s a Broadcast Band High Pass Filter from Array Solutions. (see their website for more information). This is very
useful for rejecting local broadcast stations while tuning a 160 meter antenna.
Inserting this filter will normally preserve the SWR readings around 160 meters
but the impedance and phase angle may be changed. By calibrating at the far end
of the transmission line, the AIM can
then transform the antenna data and significantly reduce the distortion caused
by the filter as well as the effect of two different types of transmission
line.
The figure above shows a scan using the standard calibration method when the composite transmission line
above is terminated with 50 ohms. This scan is highlighted
with dots along each of the three
traces: SWR, Zmag, and Phase. The SWR
shows some variation. The phase angle is off about 15 degrees at 1.95MHz.
Then a scan was done using custom calibration data obtained at the end of the transmission
line. The new data is plotted with
plain lines. Now the data is flat, indicating a nearly ideal 50 ohm termination.
The phase angle is essentially zero; it lies on the x-axis so it’s not easy
to see.
The figure above shows the results when using an LC parallel turned
circuit for a load. One scan was taken with the LC circuit connected directly to the AIM’s RF connector using standard
cal data.
Then the LC circuit was moved to the end of the composite transmission
line (shown previously) and another scan was done with the custom
cal data.
When the two scans are overlaid, they are very close to each other. This
before and after comparison was done by loading a new cal data file without
changing the scan limits. Then click Rescan and the new scan overlay the original data. The same type of comparison can also be done
by loading a stored graph file and clicking rescan.
This scan shows the effect of a High Pass Filter that has high
attention in the AM radio broadcast band. It is very effective in reducing
interference picked up on a 160 meter antenna. The filter’s cutoff frequency is
about 1.6 MHz. To cancel the effect of
the filter in the 160 meter band (1.8 to 2.0 MHz), the custom
calibration procedure can be used.
Below 1.6MHz, the cal data is noisy due to the attenuation through the
filter. This is because the instrument sees essentially the same thing when any
of the calibration loads are attached, so the compensation data is not
consistent with the specific load. Above 1.6MHz the signal through the high
pass filter is normal, with low attenuation, and the calibration is accurate.
The 50 ohm load read accurately in the frequency range of interest, 1.8 to 2
MHz.
When using new
filters or baluns that may have limited bandwidths be sure to evaluate the
results using loads with known characteristics before advancing to the antenna
itself.
Data Referred to Antenna
Sometimes it is desirable to know the impedance right at the antenna
terminals. The most accurate method
uses the Custom Calibration procedure discussed previously. In some cases it may be
impractical to disconnect the coax from the antenna feed point so another
procedure is available that only requires data from the manufacturer’s spec
sheet.
This method of transforming the data to the antenna can be used when
the properties of the transmission line (Zo, length, loss and velocity factor)
are accurately known. This does not require disconnecting the transmission line
from the antenna. It should only be used for transmission lines that employ a
single type of coax.
To select this feature, click Functions -> Refer to Antenna. Then enter the transmission line data in the dialog box. |
The length and velocity factor are combined to find the effective electrical length. Rather than trying to measure the physical length of the transmission
line and estimate its velocity factor (which may vary from one roll of coax to
another), you can measure the electrical
length directly using the AIM. Put a short circuit across the antenna
terminals (or disconnect the coax from the antenna) and scan to find the first
point where the phase angle is zero. This is listed at the top of the graph as
the first resonant freq, F1. This
frequency corresponds to one-quarter of a wavelength, so the electrical length of the line is:
Electrical_length = 0.25*(299.8/F1) meters
= 0.25*(984/F1) feet
This corresponds to the physical length of the coax divided by its
velocity factor. The electrical length is longer than the physical length since
the velocity factor is less than one.
Radio waves only care about the electrical length. When entering the cable parameters, if you
know the electrical length (by measuring F1 above), the top line can be a close
“estimate” of the physical length and the velocity factor is the electrical
length that was accurately measured, divided by this “estimate”. For example, when I installed the antenna
did I use 120 feet or 125 feet of coax?
Rather than trying to measure the length of the coax again, estimate its
length as 125 feet. Assuming the
electrical length was measured to be 185 feet, the velocity factor would be
125/185 = 0.6757. Alternatively, if you
know the velocity factor, calculate the physical length by multiplying the
velocity factor times the electrical length.
The cable loss is usually given in terms of “dB per 100 feet” on data sheets. Enter the value of loss for 1MHz, if it’s listed. If you’re interested in higher frequency
bands and want a bit more accuracy, pick a frequency close to your region of
interest and divide the loss at this frequency by the square root of
the frequency in MHz.
For example, if the loss at
10MHz is 1.5dB/100feet, enter 1.5/SquareRoot(10) = 0.47 dB/100 feet for the
loss. This attenuation value is close
to the value at 1MHz. You can enter
loss as either a positive or a negative number, it doesn’t matter.
When calculating loss, the physical length of
the cable is used. Although the
physical length of the coax may not be known with great accuracy, the loss has
a secondary effect and it’s not as critical as the electrical length. Using 120
feet or 125 feet for the loss calculation will not affect the final answer very
much, but the ratio of physical length to the velocity factor is important (as
mentioned above).
The nominal impedance, Zo, of the coax is very important. This can be found
from the manufacturer’s data sheet. It
does vary from one manufacturing run to another and it varies somewhat along
the roll of coax. This value may be fine tuned by using a known terminating
resistance of a few hundred ohms to experimentally optimize the transformed
value. After the cable is characterized by testing the actual cable or an
equivalent piece of cable (ideally from the same roll), the transformed
impedance values will be accurate to within a few percent.
Line Extension
When an adapter is used to connect a line to the AIM, there will be
some shift in the displayed parameters due to the phase shift in the
adapter. This can be compensated by
doing the regular calibration with the adapter in place. Alternatively, it can
be compensated by treating the adapter as an extension of the transmission line
and reversing the phase shift. The
extension is assumed to be a short piece of ideal coax (no loss) with a
specified length, characteristic impedance (Zo) and velocity factor. There is no limit to the length, but
typically the extension will be short. The default unit of length is centimeters.
The length can be entered in inches, feet, or meters by following the number
with an “i”, “f” or “m” (upper or lower case can be used).
The length can be positive or
negative. Enter zero to
turn the extension function off.
This graph shows the effect of inserting an adapter in series with a
piece of RG59/U coax that is terminated with a 200 ohm resistor.
A-
The first trace for the phase
crosses the frequency axis at 58.8 MHz.
B-
When the adapter, which is
6.5 cm long is inserted, the phase plot crosses at 56.5 MHz since the line
appears to be longer now and the frequency corresponding to a half wavelength
is lower.
C-
Enabling the line extension
function with length=6.5 cm, Zo=50, and Velocity Factor=1 compensates for the
extension and the last trace coincides with the first trace that was made
without the extension.
Note the value for Zo
corresponds to the adapter that is being compensated, not to the
transmission line. Depending on the
construction of the adapter, it may take some experimentation to find the
equivalent length. Only the electrical
length, not the physical length, is
important for this function, so the velocity factor can be left at 1.0 if
desired.
Rulers
and Offsets
Horizontal rulers can be defined to help visualize where a parameter
lies with respect to user specified
limits. Two rulers can be defined for:
SWR, Phase, Zmag, Rseries, Xseries, and Return Loss. These are referred to as
Ruler 1 and Ruler 2. Either one can
have the larger value.
The rulers are displayed as dashed lines with the same color as the
trace they are related to. The numeric
value is displayed at the left end of the ruler.
The ruler display can be turned on or off without affecting the value
of the rulers by clicking Rulers ON/OFF. All the rulers
can be set to zero by clicking Clear All
Rulers. This applies to both Ruler
set 1 and Ruler set 2.
Rulers for SWR (red), Zmag
(green) and Phase (magenta)
The traces can be offset by a specified amount to make it easier to see small
variations around an average value.
After the trace is offset, the Scale can be expanded to highlight
the interesting part of the trace.
Click the Scale button at the bottom of the screen to call up the Scale
factor selection dialog.
For example, the quality of a dummy load can be evaluated by offsetting
the Zmag and Rs data by 50 ohms and then expanding the vertical Zmag scale to 10 ohms full scale for better
sensitivity
The offsets can be turned off without having to change the offset
values by clicking Offsets
ON/OFF. Clicking this tab again re-enables the offsets.
All the offsets can be set to zero by clicking Clear All Offsets.
The value of the offset is displayed at the left side of each
parameter. If these labels interfere
with the some traces, the labels can be turned off while leaving the offsets in
effect. Click Labels ON/OFF.
Scan Differences
When evaluating changes to a system, it is sometimes interesting to
plot the difference between scans.
This feature makes it possible to save a scan as a reference and then
subtract it from each subsequent scan.
The scan that is presently being displayed can be saved as the
reference by clicking Difference of
Scans => Save Ref
This reference will be saved indefinitely in memory as long as the
program is active. It is cleared when the program is closed or when the
scan limits are changed. The
differencing operation can be turned on and off by clicking Enable Scan Difference. This does not affect the
reference itself. The reference scan
can be recalled for viewing at any time by clicking View Ref. The reference is not affected by this operation. The reference scan is not saved with the
init file when the program is closed.
When the differencing operation is enabled, a reminder is displayed in
the lower left corner of the graph. The difference applies to Zmag, Phase, Rs and Xs. Any of these
traces can be turned off, if desired.
New scale factors and offsets can also be used to expand the difference
traces.
This example shows how the effect of a small impedance change at the
end of a cable can be highlighted.
This shows a piece of RG58 coax 28 feet (8.5m) long, terminated with a
100 ohm resistor.
This is
saved as the Reference scan for this example.
Now we'll add a 33 pf capacitor in parallel with the resistor:
A Rescan shows the effects with both terminations
superimposed.
Displaying the difference between these scans makes it easier to see the effect of
the 33pf of stray capacitance:
The differences can be highlighted even more by plotting Zmag and Phase
on larger scales. The difference can be
displayed in the Recycle mode while tuning a circuit or antenna.
Smith Charts
Click on the button labeled “Smith” at the bottom of
the graph to open a window with a Smith chart showing a plot of the reflection
coefficient versus frequency. As the cursor is moved with the mouse over the original scan, a marker dot is displayed at the corresponding
point on the Smith chart and the relevant data is displayed on the right side
of the graph.
Note, the point that is highlighted on the Smith
chart corresponds to the frequency on the horizontal frequency axis of the
graph. This graph may be under the Smith chart and the horizontal axis, hence,
it may not be completely visible. The Smith chart window can be moved to optimize
the display.
A green circle of constant SWR
will be displayed when the SWR
Ruler #1 (see the
Setup menu) is set to a value
greater than 1.0. A second SWR ruler #2
can also be used. It will be plotted in
blue. The SWR reference values
for the rulers are shown in the left side of the Smith chart window.
A red dot marks the start of the
Smith chart trace.
The center of the Smith chart corresponds to 50 ohms
in the example below. The Zref value is specified using the Setup Menu. Zref
can be any value and it can be a complex number.
Buttons in the upper right corner of the Smith chart
window provide for zooming by factors from 1x to 5x. When the second Smith chart is displayed, it can have a
different zoom value. (see Marker
Menu)
Smith charts of the same scan with zoom values of 2x and 5x. The right hand Smith chart has an offset of 45 electrical degrees. (geometrical plot rotation is 90 degrees)
When the Smith chart has a complex trace, it may be helpful to highlight the data for each cycle around the center by plotting with different colors. This feature can be turned on and off by clicking the Colors button in the lower right hand corner of the Smith Chart.
Standard Smith Chart Color Highlighting
Reflection Coefficient in rectangular and polar coordiantes
The electrical length of the
line at a particular frequency is displayed in the lower right corner of the
Smith chart.
DSP Filters
The DSP filters are used primarily
with the PowerAIM when testing antennas that are subject to strong interfering
signals from radio stations on or near the same channel.
DSP filters 1 & 2 - A station on the same channel may cause a small but noticeable disturbance of the scan where you want to make a measurement. This interference can often be eliminated by using the DSP filters.
These filters have two parameters: center frequency & scan width. The center frequency does not have to correspond exactly to the point where the noise is centered. Also, there can be more than one burst of noise in the scan range. Typically the scan range will be small, no more than 100KHz for signals in the AM broadcast band and up to 1MHz in the FM band. The scan range is specified by its half-width. The scan will be between f1 and f2 corresponding to the center frequency minus the half-width and the center frequency plus the half-width. The numbers can be entered in kilohertz by following them with a “K” or “k”.
If 0.0 is entered for the center frequency, the DSP filter will process the data already in memory from a previous scan. This is useful for processing data saved in scan files. Otherwise, a new scan will be done with the specified scan limits and this new data will be processed.
The two filters use different techniques and one may work better that the other, depending on the situation. The calculated data will be superimposed using a heavy line over the original data so you can see its effect. The new trace can be displayed by itself by clicking on a function that causes the display to be refreshed, for example, click the scales button and then enter.
An example is shown below using DSP filter 1:
ORIGINAL SCAN: BEFORE PROCESSING WITH DSP FILTER 1:
AFTER PROCESSING WITH DSP FILTER
1:
(heavy lines are new data after
filtering)
In the example above, the heavy, highlighted lines are shown so it’s easier to see the effect of the filter. Red corresponds to SWR, orange to Rseries and yellow to Xseries.
Below are two screen shots showing
the effect of using DSP filter 2. The interfering signal in this
scan was injected using a transmitter feeding one antenna while the PowerAIM
was scanning another nearby antenna.
ORIGINAL SCAN:
AFTER PROCESSING WITH DSP FILTER
2:
TDR (Time
Domain Reflectometry)
The TDR feature enables measurement of transmission line impedances and lengths even when they are not open or shorted at the far end. This is also very useful when two or more different types of transmission line are connected in series. For example, 50 ohms and 75 ohms. The distance to the antenna can be measured without having to disconnect the line from the antenna.
For many years TDR's have been built with analog pulse generators and oscilloscopes. This requires a wideband scope (or a sampling scope) but basically the system is straightforward. This was the only practical way to implement TDR before low cost powerful computers became available.
This diagram shows the basic components of a TDR. The test pulse has a rise time that is short compared to the times to be measured. Typically the pulse has a sharp leading edge and it is much wider than any time to be measured. The trailing edge of the pulse is not used. This is called a step function. The amplitude of the pulse is not critical since the measurements involve ratios. In this figure the amplitude is 10 volts. (It could be a low amplitude pulse too.)
The output of the pulse generator is a voltage source. The leading edge of the pulse steps up to 10V at time T=0 and the voltage seen at point A is always the same regardless of the load that is connected.
There is a resistor in series with the output of the pulse generator, typically this is 50 ohms. For this example, a piece of coax is connected to the output which has a characteristic impedance, Zo, of 50 ohms. This test coax is 100 feet long and it has a velocity factor of 0.66.
The scope is connected to point
B. Keep in mind that at T=0, the coax looks
like a 50 ohm resistor to the wave that is starting to travel down it. During the initial transient stage this
value of Zo will apply until the wave has time to travel all the way to a
discontinuity in the system and be reflected back. If there is no discontinuity and the wave travels down the line
forever or is absorbed by a load resistor at the end, then the impedance at
point B will continue to be Zo forever.
In a typical case, a ideal 50
ohm coax terminated with an ideal 50 ohm resistor looks like 50 ohms at DC and
at all RF frequencies. In actual
practice, loss in the coax and skin effect complicate things somewhat but we
won't worry about that for now.
The 50 ohm resistor in series with the generator output and the Zo of the transmission line form a voltage divider of 50 ohms and 50 ohms, so one half of the voltage appears at point B at T=0.
The wave would travel down the coax at the speed of light, which is about 1 foot per nanosecond, but the dielectric material in the coax slows it down. The velocity factor is the actual velocity of the wave through the dielectric divided by its velocity if the coax did not have a dielectric (that is, the speed of light in air) . Typically this is in the range of 0.66 to 0.9, depending on the material used for the dielectric. The dielectric makes the line appear to be longer than it really is, that's why a quarter wave coax stub is shorter than a quarter wave length of the signal in air.
The voltage step in this example takes 150 nsec to travel down the coax to the open end. A scope probe at point C will see a step that is 10 volts high but it's delayed by 150 nsec with respect to the step seen at point A (the 50 ohm series resistor has negligible delay).
Wait a minute, the voltage at point B was only 5 volts and now we have 10 volts at point C. How did that happen? In fact, a scope probe placed anywhere along the coax will see the 5 volt step traveling down the line for the first 150 nsec. When it gets to the open end of the coax, the wave is reflected in phase and the 5 volt step now becomes 5 + 5 volts to form the 10 volt step seen at point C. Now, as we watch the signal travel back down the line, its amplitude is 10 volts. When the reflected wave reaches point B, the probe at point B sees 10 volts. It takes 300 nsec for the wave to travel down the line and back. The scope can measure this time which is two times the electrical length of the coax. If the velocity factor is known, then the physical length of the line can be calculated by multiplying the electrical length by the velocity factor.
The 50 ohm resistor in the generator
is the same as the Zo of the test line, so there is no reflection at the
generator and the final value seen at point B is 10 volts. It takes a short time (usually nanoseconds
or microseconds) for the voltage to reach its final value. This time delay usually isn't a factor for
ordinary radio communication . For
high speed data communication, it can be a factor when the delay is comparable
to or even greater than the time interval between data bits.
In this example, the wave at the open end of the coax was reflected in such a way that it was positive and it added to the incoming wave. The ratio of the reflected signal to the incoming signal is called the reflection coefficient. Anytime there is a change in the impedance, there will be a reflection. In this special case the reflection is equal to the incoming signal so the reflection coefficient is +1. Another interesting case is when the end of the coax is shorted. In this case the reflected signal will be inverted and it subtracts from the incoming signal so the wave that travels back toward the generator has an amplitude of 5-5=0 V. The reflection coefficient is -1. In this case, the voltage at point B will drop to zero after 300 nsec, which is what you expect since an ohmmeter at point B will read zero ohms due to the short circuit. In summary, the reflection coefficient is always in the range of +1 to -1. It can't be greater than one because the reflected wave cannot be greater than the incident wave.
Here's a practical example where the
coax is terminated in a 100 ohm resistor which does not match the
characteristic impedance of 50 ohms.
The line is said to be mismatched. To help visualize what is going to happen,
let's look at the final state of the voltage after a long time has
elapsed. The generator is outputting
10V, there is a voltage divider of 50 ohms and 100 ohms, so the final value at
point B has to be 6.66 V after things have settled down. The question is how long does it take for
this steady-state result to be reached?
In the previous examples with an open circuit and a short circuit, the reflection coefficients were +1 and -1 respectively.
The general formula for the reflection coefficient, RHO, is:
RHO = ( Zload - Zo )/( Zload + Zo )
where Zload is the load at the far end of the line and Zo is the characteristic impedance of the line. We can check this for the two extreme cases, one when Zload is infinite (open circuit):
RHO= (infinity - Zo)/(infinity + Zo) = +1
When Zload=0:
RHO=(0 - Zo)/(0 + Zo) = -1
For values of Zload between zero
and infinity, RHO varies between -1 and +1.
Therefore, for the above diagram, RHO = (100 - 50)/(100+50) = +0.33
This means that if the incoming signal has an amplitude of 5V, the reflected wave will have an amplitude of 5*0.33 = 1.66V and it will add to the incoming wave so the final value at point C at T=150 nsec is 5+1.66 = 6.66V. This reflection of 1.66 V travels back down the line and after another 150 nsec it arrives back at point B and the voltage at point B jumps up to 6.66V. Thus the final voltage at B and C is 6.66V just as we expected from the preliminary analysis when we treated the circuit as a simple voltage divider.
By measuring the reflected wave at Point B, we can work backward and calculate what Zload must be to cause this reflection. The following diagram shows the measurement point B expanded.
The reflection coefficient is the value of the second step, v2, divided by the value of the first step, v1. Note that v2 is a negative number when Zload is less than Zo.
Zload = Zo* ( 1 +
RHO)/( 1 - RHO)
Let's check some special cases
for measured values of RHO:
V1 and V2 can be measured with a scope.
If the line is open: v2 = v1
Then RHO = v1/v1 = +1
Then Zload = Zo * ( 1 + 1)/( 1 - 1) =
infinity
If the line is shorted: v2 = -v1
Then RHO = -v1/v1 = -1
Then Zload = Zo * ( 1 - 1 )/( 1 + 1 )
= 0
If the line is terminated with 100
ohms: v2 = 0.33*v1
Then RHO = 0.33v1/v1 = 0.33
Then Zload = Zo * ( 1 + 0.33)/( 1 - 0.33) = 50 * 2 = 100 ohms
If the line is terminated with 50 ohms: v2=0 (no reflection)
Then RHO = 0/v1 = 0
Then Zload = Zo * ( 1 + 0 )/ ( 1 - 0) = Zo = 50 ohms
If the line is terminated with 30
ohms: v2 = - 0.25*v1
Then RHO = -0.25v1/v1 = - 0.25
Then Zload = Zo * ( 1 - 0.25)/( 1 +
0.25) = 50 * 0.60 = 30 ohms
This is a brief introduction to the basic principles of TDR. In recent years the increased power of low cost computers has made it possible to create a TDR in software using the data from a frequency scan of a circuit. The vector network analyzer first collects data for the reflection coefficient at a number of frequencies. Then mathematical operations convert the frequency data to the corresponding time data. In this way the analyzer, which is basically a frequency domain instrument, does double duty by providing both frequency and time domain data with no extra cost in hardware.
An interesting Application Note on TDR fundamentals can be found here:
An interesting thing that can be done with a TDR is determine the impedance and
lengths of several different cables in series. This diagram shows data from an AIM for a
transmission line made up of three lengths of coax with nominal impedances of 50, 75
and 53 ohms. The composite line is terminated with 50 ohms. The numbers along the
horizontal axis are distance.
The first section of line has a measured impedance of 51 ohms and it's 5.8 feet long. The second section has a measured impedance of 74 ohms and it is (11.4 - 5.8)=5.6 feet long. The third section is (17.4 - 11.4) = 6.0 feet long.
Multiple reflections quickly complicate the interpretation of a TDR scan.
This web page has an excellent presentation on a technique for the Analysis of reflections
http://home.comcast.net/~howard.heck/3-1.ppt
This picture shows the TDR picture of a simple trap dipole antenna feed with a 53 ohm coax that is 23 feet long (indicated by the vertical cyan cursor). Since the antenna is a relatively narrow band circuit, its impedance is very high over most of the frequency range scanned by the TDR. This causes a large jump in the reflection coefficient at the far end of the coax, so it's possible to determine the length of the transmission line without disconnecting it from the antenna.
In this case the antenna is an open circuit for DC, so the impedance is very high at the feed point. If the antenna has a DC path to ground at the feed point, the TDR trace will go to zero ohms at the end of the transmission line.
The final steady state impedance displayed by the TDR is the same impedance that would be measured at DC. Depending on the horizontal scale, this final value may not be reached because of multiple reflections in the system.
The resolution in distance is approximately 0.25% of full scale for the AIMuhf and 0.5% of full scale for the AIM4170 and the PowerAIM. For distances greater than 200, the three instruments have essentially the same resolution.
TDR Operation
TDR Mode on/off - alternates between the normal Frequency mode and the TDR mode.
Linear Z Scale - If this is checked, the vertical Z display is a linear scale. If it is not checked the vertical scale shows RHO on a linear scale and Z is a function of RHO.
Show RHO Scale - If this is checked, RHO will be displayed on the vertical axis along with Z.
Lines/Dots - Plot the trace with continuous lines or dots at each measurement point.
Velocity Factor - Enter the velocity factor used for calculating distance based on electrical length. The same value is used for the whole transmission line system when multiple cables are involved. The typical value is 0.66 to 1.0. The electrical length can be displayed on the graph by setting the Velocity Factor = 1.
Meters/feet - Display distance data in meters or in feet. (meters can be selected as the default in the config file)
High Resolution - Increases the resolution along the distance axis by approximately 2x.
In the TDR mode, these functions
work similarly to the way they do in the frequency mode:
Scan - clear the screen and do a TDR scan. The AIM takes a series of impedance readings, converts these readings to the reflection coefficient at each frequency and then transforms the frequency data to time data using the inverse Fourier transform.
Rescan - do a TDR scan without erasing the screen. The color of the plot cycles through five different colors.
Scales - enter the distance and Zmag scales.
Halt - stop a scan
Quit - exit the program
The buttons at the bottom of the screen that are not used in the TDR mode are grayed out.
On the Files Menu (upper left corner of the screen), the same menu items are used for saving TDR graphs and recalling them from memory.
The AIM uses the same calibration file and config file in both the frequency mode and the TDR mode.
Setup -> Ruler 1 and Ruler 2 - Rulers for Zmag can be used.
Setup -> Average Reading - The averaging is normally set to zero (no averaging) but it can be turned on while in the TDR mode.
CURSOR DATA
This figure show how data at two points can be displayed along with their differences by moving the cursor to the first point and pressing a 1 on the keyboard, then move the cursor to the second point and press 2. This action can be repeated by moving to other points and pressing 1 or 2. Data in the small window will be updated each time. The two selected distances are indicated by blue vertical lines on the graph. Close the data window by clicking the X in the upper right corner.
The data at the cursor position is displayed in the upper right corner of the main window just like it is for the frequency mode. In this diagram, the mouse cursor is all the way to the right side of the graph at Distance=50 and the line impedance is 100.51 ohms. The value of Zo (Zref ) is used for calculating the reflection coefficient. This can entered by using the Setup menu.
AIM
Principles of Operation
The AIM4170 is based on
the same technology used in the AIM430 and AIM4160.
The AIM430 was
featured in an article in the Nov 2006 issue of QST magazine.
Block diagram
of the AIM4170:
The AIM uses two Direct Digital Synthesizer (DDS) chips. One generates the test signal and the other acts as a local oscillator to heterodyne the RF signals to the audio range. You can read about the basic principles of the DDS at this address:
http://www.analog.com/library/analogDialogue/archives/38-08/dds.pdfA 50 MHz crystal controlled oscillator drives both of the DDS chips. Inside the DDS, the clock is multiplied by a factor of 8, so the effective clock rate is 400 MHz. Program frequency resolution is a fraction of a Hertz. The output of each DDS goes to a 170 MHz low pass filter to remove the harmonics and aliases of the digitally generated signal. The output of the low pass filter is a sine wave in the range of 0.1 MHz to 170 MHz. Any amplitude variations or phase shift in the low pass filters do not affect the measurement accuracy since they affect the current and the voltage channels equally and thus cancel out when the ratio is taken.
The output of one DDS supplies the test voltage and current to the load impedance and the other DDS acts as the local oscillator to heterodyne the voltage and current signals down to 2 kHz. Audio amplifiers boost the 2 kHz signals and drive the input to the 12-bit analog to digital converter (ADC) that is inside the MSP430 microprocessor. This microprocessor is mounted inside the AIM case. The raw data is sent from the microprocessor to the external PC. The PC calculates the various data values and displays them graphically.
Reflection Coefficient:
To find the SWR (standing wave ratio) of an antenna, we first calculate the reflection coefficient. This is the ratio of the voltage that is reflected at the antenna to the voltage that arrives at the antenna from the transmitter. If all the power from the transmitter is radiated into space, there is no reflection, the reflection coefficient is zero and the SWR=1.0.
The following discussion uses the concept of complex
numbers. A tutorial on complex numbers is available in Appendix 1.
Reflection_Coefficient = Rho = ( Zload – Zo ) / ( Zload + Zo )
Zload = antenna impedance
Zo = transmission line impedance
Note that in general, Zload and Zo are complex numbers of the form:
Zload = Ra + jXa and
Zo = Ro + jXo.
Xo, which is the imaginary part of Zo, is often neglected since it is usually small compared to the real part, Ro.
Since Zload is a complex number, the reflection coefficient, Rho, is also a complex number.
The reflection of the incoming power from the transmitter is caused by a mismatch between the transmission line impedance (Zo) and the impedance of the antenna at the operating frequency. For example, if the transmission line has an impedance of 50 ohms and the antenna is a dipole with an impedance of around 75 ohms, there is a mismatch and some of the power is reflected even though the antenna itself may be very good. If the transmission line is changed to 75 ohms, the match is much better, there is less reflection and the SWR is closer to 1.0.
In the special case where the transmission line is open at the antenna (due to a broken wire), all of the power that arrives at this open circuit will be reflected back toward the transmitter and the reflection coefficient is 1.0 and the SWR= infinity. Another interesting case is when the transmission line is shorted at the antenna terminals. Again, all the power will be reflected (none is radiated) but the signal is inverted, so the reflection coefficient is now –1.0 (minus one). The magnitude is still unity (that is, +1) and the SWR=infinity.
Thus, we see that the magnitude of the reflection coefficient will be in the range of zero to 1.0 for any combination of transmission line and antenna.
The reflection coefficient also has an associated phase angle, Phase, between the incident voltage from the transmitter and the reflected voltage. The real and imaginary parts of Rho can be related to its magnitude and phase angle with the following equations:
Real_part_of_Rho = rhoa = Magnitude_of_rho * COS(Phase)
Imaginary_part_of_Rho = rhob = Magnitude_of_rho * SIN(Phase)
Rho = rhoa + j rhob
Standing Wave Ratio (SWR):
If the transmission line has no loss, the SWR is the same at all points along the line. That is, the SWR at the transmitter is the same as it is at the antenna. As the transmission line loss increases, the effect is to make the SWR measured at the transmitter appear to go down since less power is received back from the antenna. This power gets lost along the transmission line, so it does not arrive at the SWR meter and the meter responds more to the outgoing power from the transmitter. The meter thinks the antenna is a better match than it really is because there seems to be less reflected power.
The SWR only depends on the magnitude of the reflection coefficient, Rho:
SWR = [ 1 + magnitude(Rho)] / [ 1 – magnitude(Rho)]
This shows that when the magnitude of Rho = 0 (that is, the transmission line and the antenna are a perfect match), the SWR is [1+0]/[1-0] = 1 (this is the ideal case).
When the mismatch is very large and the magnitude of Rho is nearly 1, the term in the denominator approaches zero and the SWR approaches infinity.
Since only the magnitude of Rho appears in this equation, SWR is not a complex number (it’s a real number).
For calculations involving complex numbers, a very useful program can be found here:
Measurement of
Crystal Parameters
Quartz crystals can be modeled as shown below:
The series resonant frequency is the lower of the two
frequencies. It’s determined by L
and Cs. The higher parallel resonant frequency is
determined by L and Cp+Cs.
A broad scan can be done to locate the resonant
frequencies. They will be displayed at the top of the screen. Typically, these
two frequencies only differ by a few kilohertz.
Automatic
Crystal Parameter Calculation
The calculations to find the crystal parameters are
tedious so they have been combined into a procedure than can be called from
the Functions menu. First click
“Measure Crystal”. Be sure the AIM has been calibrated
with the cal loads at the point where the crystal will be connected. |
|
Then enter the series resonant frequency. This should be within 100KHz of the
series resonant frequency. When measuring an overtone crystal, enter
the appropriate harmonic frequency. |
|
After a few seconds the crystal
parameters will be displayed The data can be
saved in a file. If the file already exists, the data will be appended
(added) to it. If the file does not exist, a new one will be created with the
.xdat extension. This file is compatible with programs
that read comma separated variables (*.csv) files, like Excel.
A legend is included at the beginning of the file to show the variable
name in each column. An Excel utility can be used to sort the data to
determine the best crystals to use for a particular crystal filter. If the Parallel Capacitance is more than 20 pf it will highlighted
in red because this may
indicate there is excessive stray capacitance in parallel with the crystal
due to long leads. |
|
The following discussion goes into more
detail about the crystal calculations.
After the resonant frequencies are
located, you can change the scan limits to focus in more detail on the region
of interest.
At the series resonant frequency, the
reactances of L and Cs cancel out and the magnitude of the impedance becomes
equal to R. Therefore,
the series resistance of the crystal can be read directly at the minimum value
of Zmag on the graph; the numeric value is Rs in the Data Window.
At frequencies well below the series resonant point,
the impedance value is the total capacitance, Ctotal, which is essentially
equal to Cparallel. Cparallel is read
at a frequency equal to 78% of the series resonant frequency.
Ctotal approx = 4.5 pF
The value of Cseries is given by:
Cseries = Ctotal * 2 * (Freq_parallel
- Freq_series ) / Freq_parallel
Where Freq_parallel is the parallel resonant frequency and
Freq_series is the series
resonant frequency.
Cseries = 4.5pF * 2 *(12.12059 – 12.09557)/12.12059 = 0.0186pF
The inductance L is given by:
L = 1
/ ( 4 * pi*pi * Freq_series * Freq_series * Cs )
L =
1/(4*pi*pi*12.09557MHz^2*0.0185pf) = 9308
uH
The Q of the crystal is found by measuring the two frequencies F1 and F2 on either side of
the Freq_series where the phase angle is +/- 45 degrees.
Q =
Freq_series / ( F2 - F1 )
Q =
12.09585/(12.09591 – 12.09578) = 93045
Frequency Source
When the AIM or AIMuhf is used as a signal source, DO NOT connect it directly to the antenna connector of a transceiver. If the transmitter is accidentally turned on, the output will exceed the AIM's maximum safe input level.
The AIM can be used as a signal source for testing electronic circuits, such as radio receivers. The programmed frequency has an initial accuracy of +/-30ppm for the AIM4170 and PowerAIM. The AIMuhf will be within +/-2 ppm of the specified frequency . It can be calibrated with respect to WWV or with a counter at 10MHz. The AIMuhf has a temperature compensated oscillator (TCXO) and it will stay within 0.2ppm for several hours. The amplitude of the output is not regulated by an AGC circuit but it is flat within 1 db across any amateur band. The AIMuhf has an option to set the output amplitude over a range of 100% to 10% when using Custom Cal or the Constant Frequency mode.
The AIMuhf is limited to 200 MHz in the constant frequency mode because the output signal becomes very complex due to the presence of several aliases at high frequencies.
The nominal signal delivered to a 50-ohm load by the AIM4170 and the AIMuhf is about 30mV rms.
Enter a frequency value in the text window and click “Enter”. The frequency can then be incremented or decremented by clicking the +/- buttons.
The output frequency of the AIM can be calibrated with a frequency counter or a radio station with a known accurate frequency. The AIM has to be tuned to within 2KHz of the specified cal frequency for the calibration to be effective. The cal frequency can be any value up to 200 MHz.
For example, to calibrate the
frequency to WWV, tune in WWV at 10MHz and adjust the AIM output to zero beat
with it. Then click the “cal” button shown in the dialog box below. There is an option to enter the exact
frequency, the default value is 10 MHz. This will apply a correction factor
to all frequencies that are programmed later. The correction
is also saved in the .ini file and automatically recalled each time the program
starts up.
The value of this correction factor can be seen in the Status Window (click Help -> Status).
The frequency calibration factor can be reset to 1.00 by clicking the button called "Reset Freq Comp".
Inductance /
Capacitance Measurement
When testing components it is often convenient to have a visual display of the inductance or capacitance as a function of frequency. This shows at a glance how the parameter varies without having to read the cursor data at each frequency point. When displaying L or C the vertical scale on the left side of the graph is used. Only L or C can be plotted at one time. L and Q can be plotted at the same time.
The parameter to be plotted is selected using the Setup => Plot Parameters menu. The data is numerically the same as the cursor data displayed on the right side of the graph. The L corresponds to Xseries (Xs) and C corresponds to Xparallel (Xp). If the sign of the reactance changes during a scan, a note appears at the bottom of the graph to indicate the component is not purely inductive or capacitive over the specified frequency range.
This figure shows a family of inductance curves for coils with 1 to 4 turns on a ferrite core of type 43 material. The bottom red trace corresponds to 1 turn and the others correspond to 2, 3 and 4 turns. The value in microhenries is shown on the left axis. The magnitude of the impedance (green) and the phase angle (magenta) are also displayed. The Zmag and Phase traces are optional and they can be turned off if desired by using the Setup =>Plot Parameters menu.
This family of traces shows the inductance for 1 to 5 turns on a type 61 core. The bottom red trace corresponds to one turn and the top trace is five turns. The inductance is less than it was when using type 43 material because the permeability of type 61 is less. However, the inductance is flatter across the frequency range.
This plot shows a 100 ohm resistor in parallel with a 33 pf capacitor. The value of Rp and Cp are flat across the range. The trace labels (R, C, X) on the left are used to highlight the traces.
Appendix 1– AIMUHF
Specifications
Frequency Control: Digital Synthesizer 5 KHz - 1000 MHz
Frequency Accuracy: +/- 2
ppm (TCXO settable to WWV or a counter @ 10MHz)
Frequency program step size: 0.1 Hz min
to 10 MHz max
Calibration: software controlled (no screwdriver adjustments).
ADC resolution: 12 bits.
Measurement Ranges:
SWR: 1 to 20
Impedance: 1 ohm to 5K ohms up to 60 MHz, 2K ohm to 200MHz, 600 ohms to 1GHz
Accuracy: 1 ohm +/- 2% of reading up to 60 MHz
5% of reading up to
500 MHz
10% of reading up to
1GHz
Phase Angle: +/-90 degrees
(true phase)
Parameters displayed
include: SWR, Reflection coefficient, Q, Return loss, Magnitude of load
impedance, Phase angle of load impedance, Equivalent series resistance and
reactance, Equivalent parallel resistance and reactance. Reactance is shown as
inductance (uH) or capacitance (pF) according to the phase angle.
Two Smith Chart displays
with zoom, phase offset and markers.
Data can be referenced to
the antenna terminals.
RF Output: Programmable 0.2 to 20 microwatts (-17dBm)
max. Type-N connector standard.
Spurious output: –30dBc or better below 200 MHz
Max allowable RF input
while measuring: 140mV peak (-7dBm)
Max safe RF input: 2V peak
(+16dBm)
PC Interface: USB cable
included
Display: Graphics output
on PC screen. Mouse controlled cursor for
digital parametric readout. Markers for highlighting user specified frequencies. Audible tone and speech
output for SWR to assist in making adjustments without watching the PC
monitor.
Power Requirements: 8 to 15 VDC at 500mA max (120VAC power
supply included)
Dimensions (approx): 5” x 4” x 1.5” (12.7 x 10.2 x 3.8 cm)
Software updates are
available from W5BIG.com (no cost).
AIM4170C Specifications
Frequency Control: Digital Synthesizer 5KHz - 170 MHz ;
Stability: +/- 30 ppm
Frequency Step Size: 1 Hz to 10MHz
Calibration: software
controlled (no screwdriver adjustments).
ADC resolution: 12 bits.
Measurement Ranges:
SWR: 1 to 20
Impedance: 1 ohm to 5K ohms up to 60 MHz, 2K ohm to 170MHz
Accuracy: 1 ohm +/- 2% of reading up to 60 MHz
5% of reading up to
170 MHz
Phase Angle: +/-90 degrees
(true phase)
Parameters displayed
include: SWR, Reflection coefficient, Q, Return loss, Magnitude of load
impedance, Phase angle of load impedance, Equivalent series resistance and
reactance, Equivalent parallel resistance and reactance. Reactance is shown as
inductance (uH) or capacitance (pF) according to the phase angle.
Two Smith Chart displays
with zoom, phase offset and markers.
Data can be referenced to
the antenna terminals.
RF Output: 20 microwatts
max; UHF connector standard (UHF connector standard, N- optional)
Spurious output: –30dBc or better
Max allowable RF input
while measuring: 140mV peak (-7dBm)
Max safe RF input: 2V peak
(+16dBm)
PC Interface: RS232. Data
rate = 57.6K or 115K baud. An optional USB/RS232
adapter is available when ordering the AIM.
Display: Graphics output
on PC screen. Mouse controlled cursor for
digital parametric readout. Markers for highlighting user specified frequencies. Audible tone and speech
output for SWR to assist in making
adjustments without watching the PC monitor.
Power Requirements: 7 to 15 VDC at 250mA max (120VAC power
supply included)
Dimensions (approx): 5” x 4” x 1.5” (12.7 x 10.2 x 3.8 cm)
Software updates are
available from W5BIG.com (no cost)
PowerAIM Specifications
Frequency Control: Digital Synthesizer 0.1 - 120 MHz ;
Stability: +/- 30 ppm
Frequency Step Size: 1 Hz to 10MHz
Calibration: software
controlled (no screwdriver adjustments).
ADC resolution: 12 bits.
Measurement Ranges:
SWR: 1 to 20
Impedance: 1 ohm to 2K ohms
Accuracy: 1 ohm +/- 5% of reading up to 60 MHz
10% of reading up to
120 MHz
Phase Angle: +/-90 degrees
(true phase)
Parameters displayed
include: SWR, Reflection coefficient, Return loss, Magnitude of load impedance,
Phase angle of load impedance, Equivalent series resistance and reactance,
Equivalent parallel resistance and reactance. Reactance is shown as inductance
(uH) or capacitance (pF) according to the phase angle.
Two Smith Chart displays.
One can be phase shifted with respect to the other.
Data can be referenced to
the antenna terminals.
RF Output: 2 milliwatts
max; type-N connector standard.
Spurious output: –30dBc or better
Max stray RF input while
measuring: 25V peak.
Max safe RF input: 50V
peak.
PC Interface: RS232. Data rate = 57.6K or 115K baud. An optional USB/RS232 adapter is available when ordering
the PowerAIM .
Display: Graphics output
on PC screen. Mouse controlled cursor
for digital parametric readout. Morse code output for selected parameters. Up to 20 markers for user specified
frequencies.
An audible
tone related to SWR can be output to assist in making adjustment without
watching the PC monitor.
Power Supply: 12 VDC at 500mA max (12V battery and 120VAC
battery charger included)
Battery power source can
range from 11 to 16V. Required current is approximately 500mA when a
measurement is in process, 100mA when idle.
Auto-power-off after 10 minutes of inactivity.
Dimensions (approx): 5” x 4” x 2” (12.7 x 10.2 x 5.1cm)
Appendix 2 – Complex Numbers
A complex number has two parts: a
real part that we are accustomed to using for most everyday problems, and an
imaginary part. The imaginary part was
introduced to handle the square root of negative numbers. In ordinary circumstances, any number
squared is positive, so it seemed unreasonable for a negative number to have a
square root. This was resolved by defining a special value called “the square
root of minus one”. This is usually
symbolized by “i” in math books and by “j” in engineering
books. Using “j” avoids confusion in an
engineering context with the symbol “i” that is usually used for
current.
Complex numbers came into use about 500 years ago for solving algebraic equations, including
the familiar second order equation: ax^2 + bx + c = 0.
(note: the symbol x^2 means “the value of x squared”= x times x.)
Let’s look at a specific
example: x^2 – x – 2 = 0.
In this case the coefficients
are: a = 1, b = -1, c = -2
The solutions using the quadratic
equation are:
x = [ - b + SQRT( b*b – 4ac) ] / 2a
and
x = [ - b - SQRT( b*b – 4ac) ] / 2a
Inserting the coefficients of the
equation, we get:
x = [1 + SQRT(1 + 8)]/2 = 2
and
x = [1 – SQRT(1+8)]/2 = -1
Now, if we go back and insert x =
2 into the equation, the equation is equal to zero and we also get zero by
plugging in x = -1.
There is no problem here since we
didn’t have to worry about the square root of a negative number.
A small change of one coefficient
changes the mathematical problem considerably, as we will see now:
Let the equation be: x^2 – x + 2 = 0
a = 1, b = -1, c = +2
Changing “c” from –2 to +2 gives
us:
x = [1 + SQRT(1 – 8)]/2
and
x = [1 – SQRT(1-8)]/2
Now we have to deal the problem of
evaluating the square root of –7.
We write this as: -7 = (-1) * (+7)
Note the SQRT(A*B) =
SQRT(A)*SQRT(B), so SQRT(-7) = SQRT(-1)*SQRT(+7).
The SQRT(+7) is 2.646 and SQRT(-1)
we define as “j”, so SQRT(-7)=j*2.646.
One solution to the equation is:
x = [ 1 + j2.646 ]/2 = 0.5 +
j1.323
To confirm that the value
x=0.5+j1.323 actually does cause the equation to equal zero, we have to do some
arithmetic with complex numbers.
Addition is straightforward:
The real part of one number is
added to the real part of the second number.
Similarly, the imaginary part of
one number is added to the imaginary part of the second number.
(a + jb) + (c + jd) = (a+c) +j(b+d)
For example: ( 1 + j4 ) + ( 5 + j8 ) = (5+1) +j(4+8) =
6 + j12
Multiplication is a little tricky:
The two complex numbers have to be
multiplied term by term:
(a+jb)*(c+jd) = a*c + jb*c + a*jd
+ jd*jb
We get 4 terms. Note that j*j = -1, so the last term =
-d*b (this is a real number)
The first and fourth terms are real,
so we can add them directly to get: (a*c – d*b)
The second and third terms are
imaginary, so we can them to get:
j*(b*c + a*d)
The final result is:
(a+jb)*(c+jd) = (ac – db) + j(bc +
ad)
This is tedious. Fortunately, the
computer is good at this sort of thing, so we usually don’t have to worry about
the details.
Now we’ll finish checking our
equation by plugging in one of the answers that we found:
Let x = 0.5+j1.323
x*x = (0.5+j1.323)*(0.5+j1.323) =
- 1.50 + j1.323
Then, the whole equation = (-1.50+j1.323) - (0.5+j1.323) –2 = 0
(good)
To relate complex numbers to
electrical circuits, we make the following observations:
Resistance is a real
number.
Inductive reactance is a positive
imaginary number.
Capacitive reactance is a negative
imaginary number.
The impedance of a circuit is:
Z = R + jX, X = reactance and it can be positive
(inductor) or negative (capacitor)
For example, suppose we have a 100pF capacitor (100*10^-12 Farad) in series with a 500 ohm resistor and the frequency is 7 MHz.
At 7 MHz, the capacitive reactance
X= -1/(2*pi*7000000*100*10^-12) = -227 ohms
Note: the minus sign is
very important.
Z = 500 – j227 = impedance of the
series R-C circuit.
Real_part_of_Z =
Re(Z) = 500
Imaginary_part_of_Z =
Im(Z) = -227
The magnitude of a complex number
is the square root of the sum of the squares of the real part and the imaginary
part:
Magnitude_of_Z = SQRT( 500*500 +
227*227) = 549 ohms
The phase angle, Phase, associated
with this complex number can be calculated by:
Phase = ArcTangent(Imaginary_part
/ Real_part)
= ArcTangent(-227/500) = - 24.4 degrees
The negative angle is
characteristic of a capacitive circuit. It means the voltage is trailing
(or lagging) the current. In an
inductive circuit, the phase angle is positive since the voltage leads
the current.
Appendix 3 – Hot Keys
The following keyboard keys can be used instead of clicking buttons
with the mouse:
S – Scan (same as Scan button)
R – Rescan (same as Rescan button)
L – Enter new Limits (same as Limits button)
D – Get raw data in Point Data mode (S, Z and T can be output in Morse
code if desired).
U – Do a scan and save the data in a file named $AutoScan$.csv for
off-line processing by a program
such as Excel.
Q – Exit the program
Left Arrow – Move the cursor to the left on the graph.
Right Arrow – Move the cursor to the right on the graph.
Up or Down Arrow – Swap the last two scans. (either key can be used)
Control+UpArrow will overlay the last two scans if the scan limits are
the same.
Left mouse button – Disable the cursor movement (reverse this action
with a flag in Setup Menu.)
Right mouse button – Enter markers.
Appendix 4 – RS232/USB Operation
Because some newer computers don’t have RS232 ports, it may be
necessary to use an RS232 to USB adapter with the AIM. The AIM’s RS232 port uses only three wires.
Pin 2 is the data from the AIM to the computer, pin 3 is the data from the
computer to the AIM and pin 5 is ground.
Hardware handshaking is not used.
The signal amplitude on the I/O lines (pins 2 and 3) is approximately
+/-5V.
The comm port assigned to the USB adapter can be found
using the Windows Device Manager: Click Start à Settings àControl panel à System à Hardware à Device Manager.
If there seems to be a problem with the USB adapter, here are some tips
for checking it.
First, make sure another accessory in the computer or
another program is not tying up the comm port.
For example, if the AIM program is started two times without closing it
the first time, the second copy cannot use the comm port. This can happen when
the first AIM window gets hidden behind another window so it’s not obvious that
it is still active.
When the AIM hardware first powers up, it sends out a character string
that can be received by any terminal emulation program, such as HyperTerminal. To use
HyperTerminal, which is a standard accessory in Windows, click on the
Start button in the lower left corner of the
Windows desktop. Then, click “Programs” -> “Accessories” ->
“Communications” ->
“HyperTerminal”. Set HyperTerminal for
the following parameters:
Baud rate=57600; 8 data bits; 1
stop bit; no parity; echo typed
characters locally.
When HyperTerminal is ready, turn on the AIM power. The green LED on the AIM front panel will
turn on and the text displayed on the computer terminal is the following (or
similar):
Antenna
Analyzer AIM
from Array Solutions
Oct 13, 2006
If this is displayed, it means the
data from the AIM to the computer is okay.
If the text is not displayed, a scope or logic probe can be used
to check for pulse activity on pin 2 of the RS232 connector with the PC cable
disconnected. Pin 5 of this connector is the Ground pin. The normal voltage at pin 2 is about –5V and
it pulses to about +5V several times during a 1 second interval while the text
string is output right after the power is turned on. The PC doesn't have to be present at all,
so this output test is completely independent of the PC or the operating
system.
To check data from the computer to the AIM, type K1
(K and a one). Note the K is upper case. The command K1
will cause the AIM to turn on its Red LED.
There is no character
space between the K followed by a 1, but the time interval between typing the K and the 1 is
not critical. Then, type K0 (K followed by a zero). The Red LED should go off. This indicates the AIM is able to receive
commands from the computer; therefore, the I/O data link is working properly.
It has been found that sometimes the computer operating system will say
the USB driver is properly installed when it really isn’t. Deleting the driver and then starting over
and specifying that the driver must be reloaded from the CD may work.
A problem has also been found where Windows may not properly keep track of which comm port is in use. Some problems with the comm port can be solved by removing the port numbers that have been assigned in the past but which are no longer in use. This link has information for cleaning up the port assignments.
Check this web page for the latest information on the interface: www.w5big.com/FAQ.htm
------------------------------------------------------------------------------------------------------------
The following information on the USB interface applies to the AIMuhf, the AIM and the VNA2180.
Connect one end of the USB cable to the AIM (or VNA) and the other end to a USB port on your PC. You many need a new USB driver. (Many computers will already have the appropriate driver.) The USB interface chip used in the VNA2180 is made by FTDI (www.ftdichip.com). The chip is called FT232R and the recommended driver for several versions of Windows is 2.04.16. A link to it is on this page: http://www.ftdichip.com/Drivers/VCP.htm
If your computer is connected to the internet at the time you first plug in the USB cable, Windows XP, Vista and Win7 may be able to find the driver automatically on the internet. This may take two or three minutes.
After the driver is installed and it has assigned a port number to the AIM, use the Windows Device Manager to find the comm port number. Click "Start" in the lower left corner of the screen, then click "Control Panel", "System", "Hardware", "Device Manager". The comm port number will be shown in a list of i/o ports similar to this:
Under the AIM Setup menu tap at the top of the AIM program screen, click on Enter Comm Port and enter the port number that was assigned by Windows. This number will be saved in the setup file called AIM_xxx.ini. The default comm port number may be 6 or 5.
The usb interface can be customized for improved performance. Double click on the line that says "USB Serial Port (COMxx)". The General Properties dialog will pop-up:
Click on the tab that says "Port Settings".
The Bits per second (baud rate) does not matter.
Click on the button labeled "Advanced" (this button may be overlapping the Flow Control edit box). Click on the outer edge of the Advanced button.
The COM Port Number can be changed, if desired. Make a note of this number; this is the value that will be entered into the AIM program for the comm port. Change the parameter called "Latency Timer" from 16 to 1. This will speed up the communication.
It has been found that sometimes the computer operating system will say
the USB driver is properly installed when it really isn’t. Deleting the driver and then starting over
and specifying that the driver must be reloaded from the CD may work. Sometimes, deleting the device manager for
this port and then plugging the cable
back into the computer will cause the PC to reinstall the correct driver.
Check this web page for the latest information on the interface: www.w5big.com/FAQ.htm
The AIM programs has been used with Windows 2000, XP, Vista and
Windows 7. They have also been used
with the MAC.
Appendix 5 – Scan Data File Format
The scan data can be saved in two different formats that are suitable
for importing into spreadsheets or plotting programs.
In one file with the extension “.csv”, only the basic
data is saved in the following order:
Frequency in MHz, SWR, Rseries, Xseries, Zmag,
Phase(deg), Reflection coefficient, Return loss(dB), Percent reflected power.
The data values for one frequency are all on one
line and are comma delimited.
This file can be easily imported into a
spreadsheet for detailed analysis.
Another file with the extension “.scn” is saved with more information
about the conditions of the scan. A
detailed example is show below.
// The comments are not included in the actual file.
// Parameters included in this file: SWR, Rseries, Xseries,
// Zmagnitude, Phase
//
// If the equivalent parallel load circuit is needed:
// Rparallel=Zmag*Zmag/Rseries
// Xparallel=Zmag*Zmag/Xseries
//
09-08-05 10:57:33 // first line, date and time of the scan
110 // program version (changes
if the data format
// changes)
20 // number of data points minus 1 :
// (there are 21 data blocks below)
12 // scan start frequency (MHz)
22 // scan end frequency
0.500000000000000 // step frequency between test points
12 // freq at left side of graph plot (MHz)
22 // freq at right side of graph plot (may not be
// the same as "scan end freq"
10 // swr full scale
1000 // z magnitude full scale
100 // phase full scale (degrees)
22.500000000000000 // not used
0 // not used
0 // not used
0.500000000000000 // freq spacing between major vertical lines on graph
1 // plot_swr flag (1=true)
“comment string” // graph comment string, if any.
1010.010620117187500 // SWR : Start of data blocks; 5 items per block
// repeated 21 times in this example
0.892523407936096 // Rseries : equivalent series resistance of load
206.328903198242190 // Xseries : equivalent series reactance of load
206.330841064453120 // Zmagnitude : magnitude of load impedance
1.566470623016357 // Phase : angle of load impedance, radians
440.064605712890620 // next SWR
2.952029705047607 // next Rseries, etc
249.890869140625000
249.908309936523440
1.558983564376831
284.346221923828120 // SWR
7.125505924224854
314.254211425781250
314.334991455078120
1.548125863075256
202.912612916039060 // SWR
17.597200393676758
419.200958251953120
419.570159912109370
1.528843045234680
158.592254638671870
49.341548919677734
621.562377929687500
623.517761230468750
1.491579294204712
134.716033935546880
205.956863403320310
1158.638305664062500
1176.801147460937500
1.394876122474670
123.570953369140620
4566.038085937500000
2713.321289062500000
5311.385742187500000
0.536173224449158
114.446578979492190
421.559051513671870
-1494.078613281250000
1552.411987304687500
-1.295792102813721
108.322395324707030
105.496887207031250
-746.862121582031250
754.276245117187500
-1.430471181869507
102.702613830566410
49.549240112304688
-499.510437011718750
501.961944580078120
-1.471924185752869
97.688926696777344
29.741893768310547
-376.700836181640620
377.873107910156250
-1.492006182670593
91.975608825683594
20.779895782470703
-304.370910644531250
305.079406738281250
-1.502630472183228
88.663787841796875
15.381784439086914
-255.856430053710940
256.318389892578120
-1.510749816894531
80.219047546386719
12.850700378417969
-221.102661132812500
221.475799560546870
-1.512740731239319
76.584655761718750
10.647170066833496
-195.356231689453120
195.646148681640620
-1.516348838806152
68.956077575683594
9.652303695678711
-175.194458007812500
175.460159301757810
-1.515757203102112
65.095878601074219
8.543374061584473
-158.872634887695310
159.102188110351560
-1.517073035240173
61.215835571289063
7.745326042175293
-145.441238403320310
145.647323608398440
-1.517592549324036
58.022541046142578
7.089697837829590
-134.253402709960940
134.440475463867190
-1.518036961555481
53.652206420898437
6.746065139770508
-124.731056213378910
124.913352966308590
-1.516764044761658
50.906970977783203 // last SWR (21 data blocks in all)
6.325359821319580 // last Rseries
-116.475082397460940 // last Xseries
116.646713256835940 // last Zmagnitude
-1.516543030738831 // last phase
50 // real part of transmission line
//impedance (ohms)
0 // imaginary part of transmission line
// impedance
0 // transmission line type
0.660000026226044 //
transmission line velocity factor
0 // cable length
1 // meters or feet scale factor
1 // plot Xseries flag
1 // plot Rseries flag
0.000000045816051 //
Next five values are calibration data, not
// needed by
spreadsheet
0.000000000007380
-0.025273719802499
1.023359417915344
100.599998474121090 // last line
Appendix 6 - Battery Operation
The AIM can be powered with a battery for portable operation. Battery voltage can range from 8.0 to 15V.
The required current is 250mA (400mA for the AIMuhf) when a measurement is in
process and 50mA when idle. The power
is turned off automatically after 10 minutes of inactivity when the
Auto-Power-Off feature is enabled (this is an option in the SETUP menu at the
top of the screen).
Suitable batteries include a conventional 9V battery or a 12V car
battery. The easiest way to connect the battery is to use a barrel connector
like the wall power supply uses. This is a 5.5x2.1mm barrel connector. The
Mouser part number is: 1710-2120.
When using the AIM to test a mobile antenna on a motor vehicle, it is better to use a separate battery and not the 12V battery in the vehicle. A 9V battery will power the unit for several hours or a small 12V gel cell (sealed lead acid) battery can be used for extended operating periods. This avoids the problem of sneak paths through the ground between the DC power input and the antenna ground connection. It will also help reduce measurement noise if it’s necessary to run the engine while taking data (such as to operate the air conditioner). If it’s essential to get power for the AIM from the vehicle, be sure to put 500 ma fuses in BOTH the +12V lead and the power ground lead. A small voltage drop across the fuses will not affect the AIM since the battery voltage is much more than the required minimum operating voltage. The laptop computer being used should remain floating for the best measurement accuracy.
A 9V battery can be installed inside the AIM. There are diodes on the
card to automatically select either the wall power supply or the internal
battery. The power on/off switch will disconnect the battery too, so the
off-state current drain is only about 1 microamp.
Refer to the
Applications file under the Help menu for more
information on battery operation.
J4 is the power connector on the rear panel. The wall power supply
(or external battery) plugs in here. D1 and D11 are a cross-over circuit to
select either the external or internal power source, whichever is more
positive. The battery positive and negative connections are labeled on the pc
board. R54 is an optional resistor for trickle-charging a rechargeable
battery. This is a user-selected resistor. It is not included in the AIM but
there is a space on the PCB for mounting a through-hole resistor. |
|
This photo shows how an optional battery or battery pack can be
connected to the printed circuit board. |
|
Appendix 7 – Saving Screen Shots
Pictures of the scan can be very useful for documentation. The easiest way to paste a screen into a
program, such as Microsoft Word, Photo Shop, or
Paint Shop Pro is:
1: Press Alt-PrtScr to
copy the image to the clipboard
2: Move the cursor to the desired document or paint program and press Control-V.
A nice freeware
utility is called “PrintScreen” from Gadwin: http://www.gadwin.com/printscreen/
This utility installs very easily. It can save the whole screen, the
current window (the one with its top banner highlighted), or a rectangular area
that you can select. For saving shots
of the AIM data, the current window is probably the most useful choice. This should be saved
in gif or png format, not jpg. Gif and png formats save a better quality graphic
image than jpg. (jpg is better for
photographs.)
Newer versions of Windows have a utility called Snipping Tool that can be used to save screen shots too. Look in the
list of Accessories
files.
You can save the screen in .bmp format by clicking: File ->Save Image Bitmap
The Windows operating system also includes a standard accessory program
called “Paint”
that’s already available on all computers.
To run “Paint”, click the Start button in the lower left corner of the screen.
Then click “Programs”
Scroll to find “Accessories”
Click “Paint”
Click Control-V to
insert the screen shot into the working page.
Then save the file as a .gif , .png,
or .jpg. (.gif or .png are
preferable since the quality of the line drawing is better than it is in jpg)
Now your screen shot can be inserted into a document file or attached
to an email.
Appendix 8 – Configuration File
The configuration file (*.cfg) is used to set the screen colors and
some other display parameters. The
config file can be edited with any text editor, including Notepad. There is an editor included with the AIM
program, click Help -> Edit Config File to call it up.
A
utility called colors.exe is included with the AIM software to aid in
selecting screen colors. Move the sliders to get the desired color combination.
Note the numeric values for red, green, and blue and then enter these values in
the config file using a text editor. Several different config files can be resident in the same folder.
They are selected using the menu option: File -> Load Config
File.
The following comments are included in the config file itself.
==============================================================
This is a comment line that is displayed when the config
file is loaded from the menu.
// The user-supplied comment above will be displayed when
the config is loaded from the // menu.
// The comment line does not have to start with the //
symbols.
// There can be several config files in the same folder.
The comment in the first line is
// helpful to differentiate them.
// April 23, 2006
// AIM configuration file
// The parameters must remain in the same sequence.
// When new parameters are added, they will be placed at
the end of the file.
// Comments or blank lines can be inserted freely.
// Comments are indicated by a double-slash //
// The double-slash should start in column 1.
// If the config file is not found in the same folder as
the exe file,
// default values will be used.
// COLORS for PLOTTING:
// Colors are entered as RED, GREEN, BLUE with values 0-255
// 0=no color, 255=maximum color (max red, max green, or
max blue)
// Examples: 0,0,0
= black (no red, no green, no blue)
//
255,255,255=white (max red, max green, max blue)
//
255,0,0=bright red
0,255,0=bright green 0,0,255=bright
blue
// The included utility file: colors.exe can be used to
help select the colors.
// The numbers should be in decimal format.
// A space or comma can be used to separate the three
numbers.
// Only the first number (or set of 3 numbers for colors)
on each line is used.
// The rest of the line is a comment.
// A double-slash on a line means the whole line is a
comment.
// When experimenting with colors, the new set of colors
can be entered on the same
// line and the old ones are pushed to the right.
// The old ones will be ignored by the program but they are
handy for reference.
// There are two sets of colors for each parameter:
// The brighter color is used for the main scan (first
scan)
// A fainter shade of that color is used for the rescan
(second scan)
240,0,0 RED=SWR (or reflection coefficient) main
scan and Smith Chart trace
240,120,0 light red=SWR (or reflection coefficient)
rescan
0,200,0 Green=Zmagnitude main scan
130,240,180 light green=Zmagnitude rescan
200,0,200 Magenta=Phase main scan
200,180,240 light magenta=Phase rescan
0,200,200 Cyan=vertical cursor that is moved with
the mouse.
64,122,252 Blue=Return Loss main scan
160,160,250 light blue=Return Loss rescan
200,200,0 Yellow=Xs or Xp (whichever is selected)
main scan
255,255,0 light yellow=Xs or Xp rescan
255,174,0 Orange=Rs or Rp main scan
255,195,125 light orange=Rs or Rp rescan
250,250,200 smbackground=Smith Chart background color
230,230,230 grid color=color of gridlines on the graph
200, 0, 0
file name color (cal,config,cable_cal file names displayed on screen)
// Cursor enable:
1 = (default)
cursor is enable whether left mouse button is pressed or not.
// ( 0 means cursor
is enabled only when left mouse button is pressed. )
// Graph line width:
2 = Line width
of the plotted data (can be 1,2,3,4,5)
// Halt:
2 = Halt if right mouse button is pressed )
// ( 0 = Halt
if Halt button at bottom of screen is clicked )
//
// BAND
SELECTION: Start_freq, Stop_freq,
Step_freq (all in MHz)
// The actual
"Stop_freq" on the graph will be equal to or greater than the
// value specified
here.
// The frequencies
specified DO NOT have to correspond to ham radio bands.
// The frequencies
and the step freq can be any value within the range of the AIM.
// Example of defining a band for an AM radio station:
//
0.520, 0.620, 0.001, station=KLIF_570
// The string after "station=" is a
user supplied string to identify the band
// on the menu. It can be up to 12 characters long but it
cannot have any spaces.
// Use an underline or a dash character instead of a space.
1.7, 2.1, 0.01 // 160 meters all of
these values can be user specified.
3.4,
4.1, 0.02 // 80 meters
6.9, 7.4, 0.01 // 40 meters
13.9, 14.5, 0.01
// 20 meters
20.9, 21.6, 0.01
// 15 meters
27.9, 29.8, 0.05
// 10 meters
49.5, 54.5, 0.10
// 6 meters
143.5, 148.5,
0.10 // 2 meters
5.25, 4.25,
0.05 // 60 meters
10.0,
10.20, 0.005 // 30 meters
18.0,
18.20, 0.005 // 17 meters
24.7, 25.1,
0.01 // 12 meters
1.00, 160.00,
2.0 // Band A
1.00, 21.00, 0.2 // Band B
5.00, 45.00, 0.25
// Band C
13.00, 43.00,
0.2 // Band D
240, 240,
240 // highlight color for bands
1000 max number of data points. This can range
from 20 to 1000.
0 meters/feet (1 means use meters as default length, 0 means
use feet as default)
1000 Max delay in
milliseconds for external hardware
// (for example, Bluetooth) to respond to the RS232 data link.
//***************************************************************
// Parameters to set graph
and font sizes: (new with version 600)
// If the max width and
height are larger than the actual screen size,
// the full
screen size will be used.
// The earlier versions of
the AIM program used a max graph size of 950 x 600.
950 max graph width in
pixels (original AIM graph width was
950 pixels)
600 max graph height
in pixels (original AIM graph height was 600 pixels)
7 Default font size. 5
is very small, 6 and 7 are medium, 8 is large.
// minimum font value=3, maximum font value=10
1 Bold Data display ( 0=not bold data )
//**********************************************************************
0 Do not clear each screen during recycle
mode ( 1= do clear )
300 Delay in milliseconds while recycling. Several
seconds may be useful in some cases.
//**** versions 621 and higher:
1 RS232 baud rate 1=high, 0=low (may need to be 0 for bluetooth)
0 (1=issue cal warning if long stub is
attached while calibrating)
2.0 parameter 3
0.1 parameter 4
32 averaging for crystal’s parallel resonant
frequency measurement
1 =
don’t enable external program control,
// 2=enable but don’t show counter
// 3=enable external control and show
counter.
1 auto save graph after each scan, 0=don't save
// The next two lines are path names for the external
control files:
// The names controlfile= and outputfile= must appear first
on the line.
// The exact position on the line is not critical.
// The file and path names can be anything, they don't have
to be on the C drive.
// Don't enclose the path name in quotes.
// Use double back-slashes for folder delimiters.
// The path names CANNOT have spaces in them. Use dashes or
underscores.
// All valid filename characters are ok except spaces.
controlfile= hkey_current_user\software\AIMAIM\ : input control initiates scan
outputfile=
C:\\temp\\AutoScan.csv :output data
csv file
// Voice output
parameters:
voicefilefolder=Sounds_1\\
Folder for voice sounds (wav files)
1 num_digits=1 or 2 in fractional part for
voice output
0 0=Don't repeat the same number, 1=Repeat
the same number.
1 parameter_10 // if Zmag changes too much between data points, issue an alert:
4 = Freq resolution tolerance as a percentage
of full scale, e.g, 4=4% of full scale,
5=5%,
10=10%, etc.
// the alert can be
made less sensitive by entering a larger number, e.g., 100 will
disable it.
// The data will be valid even when the alert appears.
// END of config
file
Appendix
9 – Component Test Fixture
The AIM can be used to measure small discrete components over a specified frequency range. A convenient way to hold them is with a BNC to binding post adapter. Even though the adapter has significant stray capacitance and inductance, these stray parameters can be cancelled by the calibration procedure. The adapter shown here is from Jameco, part number 99355, which costs about $5.30 in the U.S.A. (ref: www.jameco.com) When calibrating, insert a piece of wire between the binding posts for the short circuit. Then remove the wire for the open circuit. The resistor that is used can be any 1/4W or 1/8W carbon or metal film resistor with an accurately known value. A resistor in the range of 50 to 500 ohms can be used. |
|
For testing very small surface mount components, another adapter can be made from a small piece of printed circuit board material. A cut down the middle isolates the two side. A similar cut is also made on the back side if the printed circuit board has copper on both sides. Two heavy wires (#16 - #20) can be used to connect it to the binding posts. The chip component being tested can be held securely by pressing it against the board with the eraser of a pencil or a plastic screwdriver. |
|
Here we see the surface mount adapter combined with the binding post adapter. The stray capacitance of this combination adapter will also be compensated by the calibration procedure. Capacitors in the picofarad to nanofarad range and inductors in the nanohenry to microhenry range can be measured. |
Components can be tested with a DC bias by using a
circuit similar to this:
Appendix 10 – External Program Control
In Nov 2011 a new method was developed for controlling the AIM with
functions in a DLL. These functions can
be employed in any language and they give the user full control of the data
collection, processing and
display. A separate document is
available with the title: AIM_Software_Guide.pdf. This
document (with a link to the demo program) is included in this zip package to
show how the DLL functions can be used.
A description of the original
external control method follows:
The AIM can be integrated with a test system so that measurements are
done under the control of a “Master”
program. A new procedure has been
incorporated in program version 700B which is more robust than the earlier
version.
By using this procedure, the
calibration and data processing in the AIM program are employed automatically
and the user can focus on analyzing the final data that is output to a file in
CSV (comma separated values) format.
This file can be read by Excel or the user's custom program.
The Master program communicates with the AIM by data strings in the
registry. The scan data from the AIM is
written to a CSV file in the same format that it was before.
The three parameters sent from the Master to the AIM are: Start_freq,
End_freq, and Delta_freq. The frequency
values are in megahertz. After these
parameters are written to the registry, the AIM will read them, perform the
scan and write the csv file. It then puts the name of the csv data file in the
registry so the Master knows the scan is complete and the data is ready.
To enable external control, set the flag in the config file to 2 or 3:
3 external program control:
1=disable 2=enable-don't display time, 3=enable & show time
If the flag is set to "3", a one second timer is displayed in
the upper left corner of the screen.
The names of the registry key and data file are specified in the AIM
config file:
controlfile=
hkey_current_user\software\AIMAIM\ :
input control key initiates scan (this is new in
:
AIM version 700B and later)
outputfile=
C:\\temp\\AutoScan.csv : output
data csv file
A sample file called "External_Control.eba" is included with
the zipped AIM files to illustrate the communication process. A compiled version of this sample program
can be run to get a feel for how the process works.
While the AIM is in the external control mode, the other features
operate normally. It still responds to mouse and keyboard commands.
The external control scenario
is:
1.
Turn on the AIM power and
then start the AIM program.
2.
Change focus to the Master
program.
3.
Master program writes control
parameters for setting the frequency range.
4.
AIM does a scan with the
specified scan limits.
6. Master program waits until C:\Temp\AutoScan.csv (output data
file) is created by the AIM. This data
file name can be specified in the config file.
7. Master program processes the scan data.
8.
Loop back to step 3.
This new
external control procedure only works with AIM program versions 700 and
later. It works with any model of the
AIM hardware.
Appendix 11 - Tone and Voice Output
Setup Menu:
TONES:
While in the Point
Data Tune
Mode, a tone related to
the SWR, Phase, or Zmag can be played through the PC speaker to assist in making
adjustments without watching the monitor.
When the Point Data window is open, click the button labeled Tone
On to start the
tone. (Later, click Tone
Off to stop the tone.)
The magnitude of the parameter that corresponds to the highest pitched tone can
be changed using the setup routine called “Set
High Tone Limit” under the Setup tab on the main menu at the top of the screen. For example, to make the highest tone
correspond to SWR=5, click Set High
Tone Limit and enter 5 in the dialog box.
This dialog box also appears when the parameter is first enabled. SWR is converted to a log scale for better
resolution at the low end. The Phase
and Zmag outputs use linear scales between zero (lowest tone) and the
HighToneValue (highest tone). For
Phase, the maximum limit is 90 degrees. For Zmag the maximum limit is 10K ohms.
SWR can be sounded
out verbally. Use the Setup menu to enable this function.
The tone or speech output can be played
over an extension phone or cell phone while making remote adjustments to an
antenna.
Voice Output:
The SWR value can be sounded out verbally in the Point
Data Mode. This feature is enabled
using the Setup menu -> Voice Output. This makes it
possible to make adjustments without looking at the computer monitor. It’s also possible to send the readings to
another location using telephones, cell
phones or portable radios. For example, the AIM can be located in the lab and
you can hear the SWR readings while up on a tower.
The numbers are saved in wav files. Each number is a separate file. The program breaks
up the swr reading into its separate digits and plays the appropriate wav file
for each digit. For example, the number
one is a file called: one.wav. The number two is a file called: two.wav, etc. The decimal
point is a file called point.wav. There are 11 separate files.
These are in a folder whose name is specified in the config file. The default folder is sounds_1 but the name can be anything you like. It has to be a
subfolder of the folder where the program is located. By changing the folder
name, you can have different sets of files with different voices and different
languages. The content of the wav file does not matter at all, only the name is
use by the AIM program.
There is an option in the config file to select either one or two
digits for the fractional part of the reading. Using a single fraction digit
will make the repetition rate faster.
Another config file option controls repeating the same reading. For example, if the swr reading is a
steady 1.25, you may not want to listen to this over and over. If the flag is zero, the same reading will
not be repeated twice in a row. If this flag is some value greater than zero, the same reading will be repeated every
Nth pass. For example, if the flag is equal
to 5, the same reading will be repeated after 5 measurements. Larger or smaller numbers can be used too.
This way you can tell if the system is still working, even when the reading
doesn’t change for a long time.
To get started, I made the wav files myself. This leaves a lot of room
for improvement. I’m sure some people can improve on the quality of these files
and perhaps share them with other AIM users.
Since the wav files are completely independent of the AIM program, it’s
not hard to create a new set in a new folder.
Then change the folder name in the config file. With the menu option: Help
-> Edit Config File you can switch
between voice file sets quickly. If you’re interested in an audio editor, try www.wavosaur.com.
Appendix 12 -
Band Skipping
When testing a multiband antenna, the scanning process can be speeded
up by jumping over the spaces between the bands of interest. The bands to be
scanned are defined in the config file in the same format that is already being
used. On one line, there are three
numbers: Start_freq, End_freq, Delta_freq.
These are in MHz.
The Start and End frequencies define a particular band that will be
scanned. These values do not have to
correspond to a ham band. When scanning a ham band, the entire band does not
have to be scanned. You can scan just the CW band or the phone band if you want
to.
The calibration procedure, either standard or custom cal, is not
affected at all by band skipping. The custom cal procedure (if used) is
performed at all the frequencies between the Start and End points. The only difference is that raw data points
are not read at frequencies between the selected bands during a scan.
Therefore, the whole freq range can be covered in much less time.
The Start and
End frequencies for the
scan are set by clicking the Limits button. The Step size is specified here and the delta_freq in the config file is
not used. All bands are scanned with
the same frequency step size, regardless of the value in the config file.
After setting up the Start and End frequencies, go to the Bands
menu and click on Band
Skip. This flag will alternate between on/off
each time you click it. There will be a check mark beside Band Skip when it is
enabled.
To compare the performance with and without skipping, first highlight
the bands. Then click on Scan for a regular scan. Now set Band Skip and click
Rescan. The whole scan over the bands of interest will be much faster and the
data inside the highlighted bands will be essentially the same as before, when
skipping was off.
Markers can be very helpful when adjusting multiband antennas or
filters. The SWR for each band can be displayed by analog gauges which make
it easier to visualize when the optimum
points are reached. See the Marker
section of this manual for details.
Appendix 13 - AIMuhf Expansion Port
The AIMuhf has an I/O port with two signal lines. One line is an output
driver that can be used as a clock or
a data output. The other signal line can be a data
output, data input or an analog input to the ADC. Connection to this port, J8, is made with a regular stereo plug (3.5mm).
This diagram shows the equivalent circuit of the I/O port. The ground
is isolated from the chassis. Only low currents should flow in this path. The Ring of the stereo plug is connected to the output pin which can be a clock or a data line. The functioning
of this line can change from one data cycle to the next. The other data line is connected to the tip of the stereo plug. The three functions of this line: Output,
Input, Analog Input can change for each data cycle too. Both of the lines can be used as outputs if
desired. They should be buffered in the
external accessory with a transistor or a logic gate with a low input bias
current. There are resistors in series
with these lines to protect them from external voltages up to plus or minus
12V.
R54 is only used for diagnostic tests.
The Ring signal swings from -5V to +5V. This is an output only. The current is limited by R56.
The Tip line can be digital in/out, or it can be connected to an analog
output signal (such as an op amp) in the accessory and then read by the ADC in
the AIM. This is a 12 bit ADC with an input range of 0 to 3.3V. A full-scale input voltage generates an ADC
reading of 4095.
Data Format:
Signal Line Setup Conditions:
Data 1 - output only. This
defaults to a SYNC output that pulses at the start of a scan. Functionally it is
similar to the sync output described in Appendix 14.
Data 2 - output (default), input, ADC input.
Appendix 14 - SYNC Output
On some models of the AIM and PowerAIM there is a Sync
output on the rear panel. This sync
pulse can be use to trigger a scope at the start of a scan or during a
frequency sweep.
If the Sync Out
item is checked, a pulse will be output at the start of each scan.
The AIMuhf, AIM4170C,
AIM4170B
and the PowerAIM have a sync output on the back
that generates a pulse a scan begins or when a reading is made in
the tuning mode.
The pulse from the AIMuhf
model switches between +5V and -5V. If this goes to a TTL or CMOS
circuit, the level should be clamped with a series resistor and a zener diode
to ground. Instead of a zener, the
clamping can be done with two computer diodes like the 1N4148 or 1N5811. One diode connects to ground, the other to the positive supply voltage. Alternatively, a large series resistor can
be used and the voltage will be clamped by the internal diodes of the
integrated circuit that's being driven.
Check the data sheet of the device to find the maximum value of clamp
current that is allowed. The Sync output is on the Ring terminal of the stereo jack.
The pulse
from the AIM4170B&C switches between 0 and +3V. The series impedance is 10K ohms, so the external load should
require less than 50 ua of input current.
This is a conventional RCA jack (often called a phono jack).
The sync pulse is normally at a low level. It
pulses high when the RF measurement begins. In the Scanning mode it goes low after the first
reading is complete. In the Tuning mode, it pulses high and low for every reading. In either
case the width of the pulse, Tpw, is approximately 1.5 msec times the Averaging
value. In the Tuning mode averaging
defaults to 16x so the pulse is about 25 msec wide and the repetition time is
about 350 msec. The width of the pulse
can be changed by using different values of averaging
Different sync pulse widths can be used for control purposes with an
external microprocessor or one-shot logic circuits. For example, a very wide pulse could be used to reset the
external hardware to prepare for the next series of measurements. This can be
generated by setting the averaging value to 64x or 128x.
In the Scanning mode
the sync output can be used as a scope trigger to view the frequency response
of a circuit as the frequency scans over the programmed range.
In the Tuning
mode the sync output can be used to program external loads for various tests
and the data logging feature can capture the measurement data for each test
configuration. The test configuration should be changed on the trailing edge of the sync pulse in order to give the external
circuits time to settle before the next RF reading is done. This provides more than 300 msec of settling
time between tests.
For a continuous frequency sweep, click the Sweep
Gen item on this menu. The following dialog
box pops up.
The frequency sweep goes from the Start Freq to the Stop Freq (MHz or
KHz). The difference between each frequency is the Step size. An optional settling time parameter can be
specified to let the response of narrow band filters settle. Up to four markers can be specified. These
will generate small "glitch"
pulses on the sync output which can be superimposed on the frequency sweep to
help visualize the frequency scale.
They do not have to be equally spaced.
They are specified in MHz or KHz.
After the sweep begins, the parameters are displayed in the upper left
corner of the graph. To stop the sweep, press any key or click the mouse.
The output signal amplitude of the AIM is not constant over the entire
operating range of the instrument but it is relatively flat over a narrow band.
The sync output from the AIM includes a 10K ohm series resistor. This
protects the output port and it can be used by external circuits for a
filter. The 4.3V zener is part of the
protection circuit and it does not affect normal operation of the sync output.
An optional user-supplied adapter circuit is shown that may be helpful to separate the main
sync pulse and the marker pulses to allow better triggering of the scope. This reduces the amplitude of the narrow
marker pulses but their sharp leading edge is still clearly visible. The diode is optional but it makes the
leading edge of the pulses appear sharper.
The scan start and the frequency changes are synchronized with the
leading edge of the pulses and their width is not important.
Terms and Conditions
ORDERING: visit our on-line store at www.arraysolutions.com/pricelist.htm or
phone us at (972) 203-2008 (9:00-4:00 CST) to order. More information on
ordering is shown below.
UPGRADES: Software upgrades are available at no cost.
Check the website: w5big.com for the latest version of the
program and the latest application information.
POLICIES AND
WARRANTIES:
Products Manufactured By Array Solutions - Limited
Warranty
The warranty on the AIM and related
models is two years. We warrant the functioning of the product as designed. If it fails to function as designed, we will
repair or replace it free of charge as long as you, the original purchaser, own
it and it was not damaged from lightning, other act of God or other
casualty. We reserve the right to make
that determination. We will photograph
the evidence of such non-covered damage and will be happy to show it to
you. We will listen to any reasonable
argument on your behalf to alter our opinion and do our best to reach a
mutually agreeable solution. You must
pay for shipping charges both to and from us.
Procedure for
Obtaining Warranty Service
If you have
any doubts as to whether a problem is with our product or somewhere else in
your system, please call us at our below listed telephone number. No one else besides us is authorized to
perform our warranty obligations. If
you are sure the problem is with our product, just ship it to our factory
address. Please follow the packaging
procedure explained below under "ALL REPAIRS." We will not make a warranty repair unless
you include a copy of your purchase receipt, so please include it with the
product to be repaired.
This warranty
gives you specific legal rights, and you may also have other rights that vary
from state to state.
Products Not
Manufactured By Array Solutions - Manufacturer's Warranty Only
We do not
warrant other products we sell. They
are warranted by the original manufacturer.
Please follow the manufacturer's policies and procedures for warranty
service to insure the best and quickest service.
Provisions Applicable to All Products
Very Important, retain all shipping boxes for
possible claims made to the shipper for damaged product! Not doing so will affect how your warranty
replacement is handled since we cannot replace a product that had its damage
claim rejected due to the customer not retaining the original shipping box.
In no case
will we be liable for any incidental or consequential damages, which includes,
but is not limited to, damage to equipment connected or attached to any product
we sell. Some states do not allow the
exclusion or limitation of incidental or consequential damages, so the above limitation
or exclusion may not apply to you.
Your sole and
exclusive remedy upon our failure to meet our warranty obligations to repair or
replace as described above is refund of the purchase price.
ANY
CONTROVERSY OR CLAIM ARISING OUT OF OR RELATING TO PRODUCTS WE SELL SHALL BE
SETTLED BY ARBITRATION IN ACCORDANCE WITH THE
ARBITRATION RULES OF THE AMERICAN ARBITRATION ASSOCIATION.
You may obtain warranty information at no charge by
calling us at the below telephone number or by contacting the manufacturer for
products we do not manufacture.
OTHER THAN AS STATED ABOVE, ARRAY SOLUTIONS DISCLAIMS
ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR ANY PARTICULAR PURPOSES WHATSOEVER
AND ANY IMPLIED WARRANTIES.
ALL REPAIRS:
When sending in a product for service, please carefully
"double" box it and ship it insured for your protection. Please include a note clearly describing the
problem, how you wish the item returned to you and how you wish to pay for the
service if it is not covered by our warranty.
Package your item properly.
Array Solutions is not responsible for merchandise damaged in
shipment. Our service rate is $50 per
hour (2 hr. minimum).
NON-REPAIR RETURNS:
Returned items must be received by us in original, as-shipped, condition
including the original box, manuals, accessories, and copy of sales
receipt. Returns must be within 14 days
of purchase. Returned items are subject
to a 15% restocking fee. Shipping
charges are not refundable.
Phone
and FAX:
Tel:
972 203 2008
Fax: 972 203 8811
Factory Address: Array Solutions 350 Gloria Rd. Sunnyvale TX 75182
Sales Outside
the United States: Purchase of products
by those outside the United States of America constitutes purchaser's agreement
that all questions concerning the respective rights and obligations arising
under the terms of purchase are subject to arbitration under the provisions of
Texas Civil Practice and Remedies Code Section 172.001 et seq.
On Line: Ordered on line. Our secure, dedicated
server is one of the fastest and protected with 128-bit encryption. You can be
sure of your privacy.
PAYMENT: We accept Visa, Mastercard, money orders, cashiers checks and personal
checks, and bank transfers. Phone order customers please be ready to provide
card number, expiration, issuing bank, billing and shipping address plus home
& work phone (e-mail customers please provide this information also). We
request this information for your protection. Money orders, cashiers checks and
personal checks may require 2 weeks to clear from un-established customers
and/or large purchases.
MINIMUM ORDER: We have no minimum. Your order is
important to us whether it is for $5 or $5000.
SHIPPING of Array Solutions Stocked Products: We ship
most orders within 5 business days (UPS Next Day and Second Day Air orders are
shipped within 24 hours). We ship Monday through Friday. Our normal method of
shipment is UPS Surface International
orders are sent with Air mail parcel post or EMS. We do not ship C.O.D. Array
Solutions reserves the right at any time after receipt of your order to accept
or decline your order for any reason. You may request shipment by other means
such as FedEx, UPS Next Day Air, UPS Second Day Air or Air Freight for export
orders. Equipment is packed very carefully and sent insured. Please provide us
with a full shipping address. UPS cannot ship to a Post Office box.
Shipping of ICE Products is usually shipped to you
Factory Direct from ICE. ICE ship dates vary extensively due to customer demand
and products in stock at their facility. We cannot be held liable for the ship
date of their products, but we will try to expedite the order if we are told by
you that you need it by a certain date. Please order these products early.
PRICES: Prices and specifications are subject to
change! We are not responsible for typographical errors.
SALES TAX: Array Solutions shall automatically charge
and withhold the applicable sales tax (8.25%) for orders to be delivered to
addresses within the state of Texas. Each customer shall be solely responsible
for all sales taxes, or other taxes, on orders shipped outside Texas.
FOREIGN ORDERS: Foreign orders must be payable in
U.S. dollars. You may use Visa, Mastercard, postal money order or bank check
drawn on a U.S. bank with transit numbers imprinted on bottom of the check or
bank transfer to our account. We will supply you our banking information upon
request. The customer is responsible for shipping on warranty service. We do
our best to deliver to you using the more expensive EMS (Express Mail Service)
parcel post which allows us a tracking number to follow up if there is a
problem. We have experienced some long deliveries to certain countries like
Spain, Portugal, Italy and some South American countries. We may insist on
using a shipping service such as UPS, DHL, FEDEX, or Freight Forwarder for
certain countries or large orders due to past losses in your country.
SHIPPING TO Array Solutions: When you ship your
equipment for service it is important that it is packaged properly. Use the
original box and packing material if possible. If you do not have the original
box, wrap the equipment well with bubble wrap then place this in a box with
adequate packing material. All equipment should then be placed in a second box
with packing material to help ensure its safe arrival to us. Please be sure to
insure your package for its value. Array Solutions will not be responsible for
equipment that is damaged in shipment to us.
CUSTOMER INFORMATION: Array Solutions does not sell,
rent, lease or trade customer information with any other company.
LIMITATIONS OF LIABILITY: Except as expressly stated
herein, Array Solutions. makes no representations or warranties, either express
or implied, of any kind with respect to products sold on the Array Solutions
website, including but not limited to, merchantability and fitness for a
particular purpose. You agree that the sole and exclusive maximum liability
that Array Solutions has arising from any product sold on the Array Solutions
web site shall be the price of the product ordered. In no event shall Array
Solutions, its directors, officers, employees or other representatives be held
liable for special, indirect, consequential, or punitive damages related to
products sold.
RETURNS: Returned items must be received in the
original condition including the program CD, power supply, cable, and accessory
connectors. Returns must be within fourteen days of receipt. If the original box is not available,
suitable boxes can be obtained from the post office without charge. Use the
original packing materials if available or the equivalent. For maximum
protection of your equipment, insure the package and request a signature when
it is received. Shipping cost is not
refundable. We will pay the return shipping for warranty repairs.
WARRANTY SERVICE: Products have a one year limited
warranty. This includes parts and labor. We will repair or replace the unit at
our discretion.
CIRCUITS
AND PROCEDURES used by this vector impedance measurement system are covered
by one or more patents pending. |